Rocket 4D Evaporation System
User Manual
46
04-7848 Issue 2-4 August 2014
Priming the cooling system
To prime the cooling system, follow the instructions in the chiller operating manual.
Note:
T
he following additional steps may be required if the Rocket is positioned a
significant height above the chiller.
1.
Switch off the chiller.
2.
Open the reservoir filler on top of the chiller.
3.
Disconnect the coolant hose from the evaporator
coolant in
connector.
4.
Hold the end of the hose over the chiller reservoir filler (so any coolant expelled from
the hose will drain back in to the chiller) and perform the priming procedure as
instructed in the chiller operating manual.
5.
When coolant starts to flow from the hose, switch off the chiller.
6.
Reconnect the hose to the evaporator
coolant in
connector and perform the priming
procedure as instructed in the chiller operating manual.