Rocket 4D Evaporation System
User Manual
16
04-7848 Issue 2-4 August 2014
Method guide
The Rocket 4D can store up to ten individual methods which can be accessed from the
Method Select
menu. Factory loaded methods are described below (methods on your
Rocket may differ if they have been over-written).
Factory loaded methods
Feed enabled system
Number Method
Mode
Solvent group
01 Feed Low BP
Feed
Single solvent or
Mixture of solvents with boiling points (within 20°C)
02 Feed Medium BP
Feed
Solvent boiling point > 90°C (including DMF)
03 Feed Aqueous
Feed
Water
04 Feed HPLC
Feed
Mixtures of solvents with dissimilar boiling points (more
than 20°C difference)
05 Batch Low BP
Batch
Single solvent with boiling point below 90°C
Mixed solvents: all boiling points within 20°C and below
90°
06 Batch Medium BP
Batch
Solvent boiling point > 90°C
07 Batch Aqueous
Batch
Water based solutions where freezing at 0°C is
anticipated
08 Batch HPLC
Batch
Mixtures of solvents with dissimilar boiling points (more
than 20°C difference)
09 No Method Loaded
10 No Method Loaded
Non-feed-enabled system
Number Method
Solvent group
01 Batch Low BP
Single solvent with boiling point below 90°C
Mixed solvents: all boiling points within 20°C and below
90°
02 Batch Medium BP
Solvent boiling point > 90°C
03 Batch Aqueous
Water based solutions where freezing at 0°C is
anticipated
04 Batch HPLC
Mixtures of solvents with dissimilar boiling points (more
than 20°C difference)
05 No Method Loaded
06 No Method Loaded
07 No Method Loaded
08 No Method Loaded
09 No Method Loaded
When custom methods are loaded, unused method slots may be used. Alternatively the
factory pre-loaded methods may be over-written.