CURRENT LOOP INSTALLATION
WIRING CONNECTIONS
When wiring the 20 mA current loop, remove the 7-position terminal block
(TBD), located on the right side of the top board. Refer to the numbers listed with
the terminal descriptions below or on the top label, and install each wire in its
proper location on the terminal block. When all connections are made, replace
the terminal block into its proper location.
It is recommended that shielded (screened) cable be used for serial
communications. This unit meets the EMC specifications using ALPHA #2404
cable or equivalent. There are higher grades of shielded cable, such as, four
conductor twisted pair, that offer an even higher degree of noise immunity.
TERMINAL DESCRIPTIONS FOR TERMINAL BLOCK TBD
1. -20 mA SRC (COMM.) - Common
2. PRINT REQUEST - The Print Request Terminal is pulled low to request the
Gemini to transmit according to the Print Options mode that has been selected.
(Minimum Activation time = 25 msec.)
3. +20 mA SRC - The 20 mA SRC terminal provides the source current for one of
the loops.
4. SO- (Serial Out-) -
5. SO+ (Serial Out+) -
The Gemini transmits the requested data on these terminals. They are
connected in series to the receive input of the device to be connected.
6. SI- (Serial In-) -
7. SI+ (Serial In+) -
The Gemini receives commands on these terminals. They are connected in
series with the transmit or output terminals of the device to be connected.
Note: The serial Input terminals must be held in the mark condition (current on)
in order for the Gemini to respond to a Print Request terminal activation.
SERIAL DIP SWITCH SET-UP
The Serial DIP switches are accessible through the side of the Gemini. A list of
the DIP switch positions and their functions are shown in the figure at right.
BR0 & BR1, BAUD RATE - Set-up is shown , at right. When changing the Baud
Rate, the unit should be powered-down and then powered back up again. The
unit will only recognize a baud rate change upon power-up, after activating the
“Print Request”
terminal or after a few characters have been sent at the new
baud rate
(If the two previous conditions have not occurred, the Gemini will
see the characters as erroneous and it will check the baud rate and set itself to
operate at the new rate).
PR.ID - PRINT ID. - When this switch is in the up position, the Gemini will print
the unit address, data value ID and the data value when a transmission is
requested. The unit will also insert a 400 msec delay between transmissions
when the
“P”
command or Print Request terminal is used. This switch
position is generally used when the unit is connected with a printer.
When the switch is in the down position, the Gemini will transmit only the data
value, without the unit address and data ID. The 400 msec delay, described above,
will not be inserted. This switch position usage is intended for applications where
the Gemini is communicating with a computer. In these circumstances printing
the address and value ID and inserting a 400 msec print delay is usually
unnecessary and needlessly slows down communication throughput.
PC0, PC1, & PC2, PRINT OPTIONS - Used to control which values are
printed out when the Print Request terminal is activated or when the Transmit
per Print Options command “P” is sent to the Gemini 4000.
AD0, AD1, AD2, & AD3 UNIT ADDRESS - These switches are used to give
each unit a separate address when more than one unit is connected in the Loop.
See below for Switch Set-up.