C57 HPW
11.2 Removing the diaphragm
1. Move bonnet
A
to the closed position.
2. Unscrew diaphragm
2
.
2
A
3. Clean all parts of the remains of product
and contamination. Do not scratch or
damage parts during cleaning!
4. Check all parts for potential damage.
5. Replace damaged parts (only use
genuine parts from GEMÜ).
11.3 Mounting the diaphragm
11.3.1 General information
Important:
Mount the correct diaphragm that
suits the diaphragm globe valve
(suitable for medium, medium
concentration, temperature and
pressure). The diaphragm is a
wearing part. Check the technical
condition and function of the
diaphragm globe valve before
commissioning and during the
whole term of use. Carry out
checks regularly and determine the
check intervals in accordance with
the conditions of use and / or the
regulatory codes and provisions
applicable for this application.
Important:
Incorrectly mounted diaphragm
may cause diaphragm globe valve
leakage / emission of medium. In
this case remove the diaphragm,
check the complete diaphragm
globe valve and diaphragm and
reassemble again proceeding as
described above.
32 / 44