Environmental conditions
Data
Þáttur
Temperature: +15 to +30°C
Operating conditions
Humidity: 30 to 70% (no dew condensation)
Temperature: -10 to +40°C
Non-operating conditi-
ons
Humidity: 20 to 70% (no dew condensation)
Temperature: -25 to +70°C
Transportation &
storage conditions (wit-
hin 96 hours)
Humidity: 10 to 80% (no dew condensation)
151 W/h (reader block + FLA Image Eraser)
Heat radiation
It is recommended to lower the lighting level to about 20
lux when moving a sample from the cassette onto the
Phosphor stage after exposure. 20 lux is the brightness
level at which most people can read newspaper-sized text
after he/she becomes accustomed to the brightness.
Lighting
Aðeins til nota innandyra
Location of use
2,000 m above sea level
Maximum altitude for
use
Transient overvoltage category II
Yfirspennuflokkur
Pollution Degree 2
Applicable Pollution
Rating
Noise levels
Data
Þáttur
70 dB (A) or lower
Suð
IP20
Degrees of protection
provided by enclosure
Getting Started with Typhoon FLA 7000 28-9607-64 AD
115