á bls. C
Reglulegt viðhald
Cleaning the outside of Typhoon
FLA 7000
Clean the outside with a moist soft cloth and a mild detergent. Wipe afterwards with a
dry soft cloth.
Maintenance of the SHG laser
If you use the SHG laser in the Typhoon FLA 7000, it will require periodical calibration.
When the Typhoon FLA 7000 is switched on, it automatically executes calibration. You
do not need to do a manual calibration if the Typhoon FLA 7000 is used at least once a
month.
AÐVÖRUN
Activate the Typhoon FLA 7000 at least once every 30 days to ex-
ecute calibration. This calibration makes sure that the SHG laser
operates properly. Failure to calibrate the Typhoon FLA 7000 every
30 days reduces the lifespan of the SHG laser.
Operation procedures:
1
Turn on the Typhoon FLA 7000 and the computer.
2
Wait until warm-up and self-diagnosis is completed, and the scanner is ready. Only
the power lamp on the upper left panel is lit. The automatic calibration has now been
performed.
3
Turn off the Typhoon FLA 7000 and the computer.
Replacing the filters in the
Typhoon FLA 7000
See the instruction in
á bls. D Installing and replacing filters, á bls. 106
Getting Started with Typhoon FLA 7000 28-9607-64 AD
105
C Reglulegt viðhald