When there's too much leakage current …
Vivid T9/Vivid T8 – Basic Service Manual
10-31
5795591-100 English Rev. 3
10-8-3 Probe Fails
Test the probe in another connector to isolate if the fault lies with
the probe or the Ultrasound system. Or Change another probe
to confirm if the fail is caused by console.
NOTE:
Each probe will have some amount of leakage, dependent on its
design. Small variations in probe leakage currents are normal
from probe to probe. Other variations will result from differences
in line voltage and test lead placement. The maximum allowable
leakage current for body surface contact probe differs from
inter-cavity probe. Be sure to enter the correct probe type in the
appropriate space on the check list.
If excessive leakage current is slot dependent, inspect the
system connector for bent pins, poor connections, and ground
continuity.
If the problem remains with the probe, replace the probe.
10-8-4 Peripheral Fails
Tighten all grounds. Ensure star washers are under all ground
studs.
Inspect wiring for bad crimps, poor connections, or damage.
10-8-5 Still Fails
If all else fails, begin isolation by removing the probes, external
peripherals, then the on board ones, one at a time while
monitoring the leakage current measurement.
10-8-6 New Unit
If the leakage current measurement tests fail on a new
Ultrasound system and if situation can not be corrected, submit
a Safety Failure Report to document the Ultrasound system
problem. Remove Ultrasound system from operation.
10-8-7 ECG Fails
Inspect cables for damage or poor connections.
Summary of Contents for T8
Page 5: ...Vivid T9 Vivid T8 Basic Service Manual i 3 5795591 100 English Rev 3 ...
Page 6: ...i 4 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 7: ...Vivid T9 Vivid T8 Basic Service Manual i 5 5795591 100 English Rev 3 ...
Page 8: ...i 6 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 9: ...Vivid T9 Vivid T8 Basic Service Manual i 7 5795591 100 English Rev 3 ...
Page 10: ...i 8 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 11: ...Vivid T9 Vivid T8 Basic Service Manual i 9 5795591 100 English Rev 3 ...
Page 12: ...i 10 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 13: ...Vivid T9 Vivid T8 Basic Service Manual i 11 5795591 100 English Rev 3 ...
Page 14: ...i 12 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 15: ...Vivid T9 Vivid T8 Basic Service Manual i 13 5795591 100 English Rev 3 ...
Page 20: ...i 18 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 28: ...i 26 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 158: ...System Setup 3 78 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 210: ...Service Adjustments 6 4 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 308: ...Renewal Parts 9 12 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 344: ...Care and Maintenance 10 36 Vivid T9 Vivid T8 Basic Service Manual 5795591 100 English Rev 3 ...
Page 349: ...GE ...