![GE T2100-ST1 Operator'S Manual Download Page 37](http://html.mh-extra.com/html/ge/t2100-st1/t2100-st1_operators-manual_132902037.webp)
Operating Instructions
2097937-001 Rev E
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
37
12 June 2017
Pull Tether Switch Check
NOTE:
Verify proper operation of the pull tether switch assembly
every month.
With the belt moving at a relatively high speed,
pull the wrist strap to activate. The treadmill will
have a controlled stop to 0.0 speeds. The running
belt will have resistance preventing the free
movement of the running surface. To reattached
the pull tether, attach clip to the original position
on the switch. The treadmill will return to 0.0%
elevation.
Use the controlling equipment to terminate the
exercise session and turn off the treadmill.
Adjustable Handrail Option
The adjustable handrail is an optional feature that allows you to
change its position to accommodate patients of various heights. If
this option is installed on the T2100-ST series treadmill, use the
following procedure to adjust its position.
CAUTION
POTENTIAL BODILY INJURY
–
Attempting to reposition the
adjustable handrail while the running belt is operating could
result in bodily injury.
Make sure the running belt is off before performing the
following procedure.
1.
Turn the adjustment knob counterclockwise twice to unlock
the handrail.
2.
Pull out the adjustment knob to release the handrail.
3.
Adjust the handrail to the desired height.
4.
Release the adjustment knob to secure the handrail in the
current position.