Product Overview
26
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
2097937-001 Rev E
12 June 2017
110-120 VAC
NEMA 5-20R
Single Phase
220/240 VAC
NEMA 6-15R
Split Phase
230 VAC
AS/NZS 3112
220/240 VAC
CEE 7/7 EURO
250 VAC
Type N BRAZIL
230 VAC
BS1363
240 VAC
BS546 3 PIN
250 VAC
Type K DANISH
240 VAC
GB 1002
250 VAC
Type H ISREAL
250 VAC
Type L ITALY
250 VAC
Type J SWISS
WARNING:
The T2100-ST1 and T2100-ST2 treadmills must be grounded
to reduce the risk of electrical shock. If a malfunction occurs,
grounding provides a path of least resistance for an electric
current. Ungrounded connections must not be used.
No other equipment may be used on the electrical circuit with
the treadmills. Do not use extension cords. Using a shared or
unreliable circuit can also cause the treadmills to
unexpectedly shut off, potentially resulting in injury to the
patient.
Ensure the master power switch is in the off position before
plugging in the T2100-ST1 or T2100-ST2. A power surge could
damage the sophisticated electronic system of the treadmills.
NOTE:
The T2100-ST1 and T2100-ST2 Treadmills must have their
own dedicated power outlet.