![GE T2100-ST1 Operator'S Manual Download Page 31](http://html.mh-extra.com/html/ge/t2100-st1/t2100-st1_operators-manual_132902031.webp)
Assembly and Setup
2097937-001 Rev E
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
31
12 June 2017
machine to another location. (
See “
Running Belt Tension
Adjustment
” on page
41 for details.)
Final Setup - Running Belt Tension Adjustment
Your T2100
®
running belt has been pre-tensioned at the factory and
run for 16 hours prior to shipment. It may, however, be necessary to
adjust the belt tension when the treadmill is run in its final location.
A loose belt tends to hesitate or stick with a heavy foot plant. If your
belt needs tensioning, the adjustment procedure can be found in
“
Running Belt Tension Adjustment
” on page
42.
NOTE:
Improper adjustment could cause the treadmill to hesitate
and cause a trip and fall hazard. Because this adjustment is
not covered under your warranty, it is important that you
review these instructions thoroughly before proceeding.
Final Setup - Drive Belt Tension Adjustment
The drive belt tension has been pre-set at the factory to minimize
maintenance. If there are indications that the drive belt has
stretched and become loose, refer to "Drive Belt Tension
Adjustment" in the Preventive Maintenance chapter of the T2100-
ST1 / T2100-ST2 Field Service Manual. Symptoms of a stretched
drive belt could include increased noise.
Final Setup - Test Plug Procedure
Each T2100-ST1 and T2100-ST2 treadmill includes an RS-232 test
plug that enables you to test the operation of the treadmill without
the ECG unit attached. The plug is located on the left side of the
treadmill secured to the frame by Velcro
®
. The plug is to be used
only for testing the treadmill. Do not stand on or use the treadmill
while testing.
Item
Description
1
Test Plug Location on Left
2
Circuit Breaker Array
1
2
3
4