![GE T2100-ST1 Operator'S Manual Download Page 25](http://html.mh-extra.com/html/ge/t2100-st1/t2100-st1_operators-manual_132902025.webp)
Product Overview
2097937-001 Rev E
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
25
12 June 2017
Communication Ports
RS232 Female Serial port
USB 1.0 “B”
port
Floor Surface Footprint
33.0 in x 78.5 in, 84cm x 200cm level surface. (
See “
Location
” on
page 29.)
Operating and Storage Condition Recommendations
Operating Temperature Range: 4.5
to +38
C (+40
to +85
F)
Storage Temperature Range: -40
to +70
C (-40
to +158
F)
Operating and Storage Relative Humidity Range: 10% - 90%,
non-condensing
Altitude: -50 to 9,842 feet (-15m to 3000m); derate performance
by 5% per each additional 500 feet (152m) above 5,280 feet
(152m)
Power Requirements
The T2100-ST1 is designed to operate on a dedicated 110-120 VAC
20-amp power supply and the T2100-ST2 is designed to operate on
a dedicated 220/240 VAC 15-amp. Make sure that the treadmill is
connected to an outlet that looks like the following illustration.
This product is equipped with a three-wire grounding-type plug. The
plug will only fit into a grounding-type outlet. This safety feature
must not be disabled. Contact a qualified electrician if you are
unable to insert the plug into your outlet, or uncertain if the outlet
meets local electrical codes. Polarized outlets such as NEMA 5-20
and CEE7/7 must be verified for proper polarity configuration before
plugging in the T2100-ST2. Incorrect polarization of the outlet could
cause failure of onboard electrical components or cause electrical
shock. Proper grounding is necessary for the equipment to meet
acceptable current leakage standards consistent with the
standards to which it was certified.