T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
35
12 June 2017
4
Operating Instructions
Before operating the T2100-ST1 or T2100-ST2, familiarize
yourself with the following control safety features of the
treadmills.
Electrical Safety Tests
The electrical safety of this installation is the responsibility of
the customer, not GE Medical Systems
Information
Technologies
. In hospitals, contact your in-house biomedical
technician, electrician, or technically qualified personnel.
Outside of hospital, contact your hospital affiliation of these
services. Otherwise, contact GE Medical Systems
Information
Technologies and open a customer-billable service call.
Before using the treadmill, have qualified personnel perform
the test listed below:
AC line voltage test to verify the power outlet is
properly wired.
Ground continuity test to verify all exposed metal is
properly grounded.
Leakage test to verify the equipment passes all
applicable leakage tests.
Your in-house biomedical technician, electrician, or
technically qualified personnel can find instruction for
performing these tests in the T2100-ST1/T2100-ST2 Treadmill
Service Manual shipped with your system.
Operating Controls
The treadmill has three operating controls: the power switch,
emergency stop switch, and pull tether. The power switch is located
on the rear panel, the emergency stop is located on the right hand
rail, and the pull tether is located on the left handrail.