2 Installation
K0443 Revision A
2 - 6
[EN] English
Note:
(Ref: Section 6, Reference and Specification)
for details of other system components.
Figure 2-4, Pneumatic Connections with vacuum supply
Note:
The PACE option IO-VAC-SYS Vacuum System Check Valve Kit should be used in the
vacuum line mounted near the PACE CM -ve port to exhaust most HP gas directly to
atmosphere. The vacuum buffer volume needs to be rated at least to the highest
system pressure.
Note:
(Ref: Section 6,
Reference and Specification)
for details of other system components
1) Pressure source
2) Conditioner
3) Filter
4) Regulate to between 110% full-scale and MWP
5) Diffuser*
6) Unit under test
7) Optional reservoir
†
8) Protection device
9) Optional differential connection
10) Oil mist trap
11) Vacuum source
12) Normally open electrical release valve
13) Check valve **
14) Manual external vent valves
a) Atmosphere
1
2
3
4
5
6
7
8
9
13
7
10
12
11
a
OUTPUT
+
-
SUPPLY
VENT REF
14
14