6 Reference and Specification
K0443 Revision A
6 - 30
[EN] English
Example Program
Note:
Changes to instrument settings made in a test program remain valid only for the test
program.
The instrument reverts to the pre-test settings on completion.
Step Command
Argument Action
START
Program start
1
UNITS
mbar
Select units, mbar
2
RATE
100
Select rate, 100 mbar/min
3
IN LIMITS TIME
10 (00:00:10) secs
4
IN LIMITS
Set In Limits Band
5
RESOLUTION
5
Display resolution, 5 digits
6
SETTLING
No overshoot
7
TEXT
Operator instruction, e.g.”Connect UUT”
8
ZERO
9
SET-POINT
400
Set-point, 400 mbar
10
CONTROL
Controller ON
11
WAIT IN LIMITS
Wait for In Limits Condition
12
BEEP
Beep on, approx. 1 sec, Beep off
13
MEASURE
Switch to Measure (controller off)
14
DWELL
30
Wait for 30 sec
(00:00:30)
15
SET-POINT
800
Set-point, 800 mbar
16
CONTROL
Controller on
17
WAIT IN LIMITS
Wait for In Limits Condition
18
BEEP ON
Beep on, approx. 1sec, Beep off
19
MEASURE
Switch to Measure (controller off)
20
TEXT
Operator instruction, e.g.
(Wait for beep, record pressure)
21
DWELL
30
Wait for 30 sec
22
BEEP
Beep on, approx. 1 sec, Beep off
23
TEXT
Operator instruction, e.g. “
Min pressure allowed 785 mbar”
24
PAUSE
WAIT,
(for operator input to touch single step)
25
VENT
Vent
END
Program end
Summary of Contents for PACE5000
Page 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Page 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Page 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Page 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...