PACE Pressure Controller User Manual
[EN] English
6 - 31
K0443 Revision A
Programming Loops
To program a loop, use the GOTO command.
Include the COUNT command in the loop for counting the number of loop cycles.
Note:
The test program commands do not include tests for conditional jumps.
To stop a test program from looping, STOP must be selected by the operator.
Example of programming a loop
Step Command
Argument Action
START
Program start
1
UNITS
mbar
Select units, mbar
2
RATE_VALUE 100
Select rate 100mbar/min
3
RESOLUTION 5
Display resolution 5 digits
4
IN LIMITS
Sets the In-limits band
5
IN LIMITS TIME
10 (00:00:10) secs
6
SETTLING_N_O zero
No overshoot
7
TEXT
Operator instruction, e.g.“Connect UUT”
8
ZERO
Performs a sensor zero
9
SET-POINT
400
Set-point, 400 mbar
10
CONTROL
Controller ON
11
WAIT IN LIMITS
Wait for In-limits condition
12
BEEP ON
Beep on, approx. 1sec, Beep off
13
MEASURE
Switch to measure (controller off)
14
DWELL
30
Wait, 30 sec
15
SET-POINT
800
S
et-point, 800 mbar
16
CONTROL
Controller on
17
WAIT IN LIMITS
Wait for In-limits condition
18
BEEP ON
Beep on, approx. 1sec, Beep off
19
MEASURE
Switch to measure, controller off
20
COUNT
Increment loop counter
21
VENT
Vent
22
GOTO
9
Loop back to program line 9
END
Program end
Summary of Contents for PACE5000
Page 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Page 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Page 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Page 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...