PACE Pressure Controller User Manual
[EN] English
2 - 9
K0443 Revision A
Procedure
1.
Locate bracket in rack assembly
.
2.
Remove the four M3x10mm countersunk screws from each of the instrument side
panels.
3.
Locate the two brackets
on each side of the instrument.
4.
Secure with the four countersunk screws.
5.
Support the instrument and connect the cables and pipes.
6.
Refer to the electrical connections below before fitting the instrument into the rack.
7.
Temporarily locate the two spigots
*
to each side of the equipment rack.
8.
Locate and slide the instrument into the rack.
9.
Locate the instrument on the spigots
*
.
10.
Secure the instrument in the equipment rack with two of the screws and washers
(supplied).
11.
Remove the two spigots
*
and replace with the remaining two screws and washers
(supplied).
2.6
Electrical connections
WARNINGS
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
Connecting
(Ref: Figure 2.7 Electrical connections)
1.
Install an accessible power isolator to use as the disconnecting device in the power
supply circuit.
2.
For the power input power supply range, Installation Category and VA
(Ref:
Introduction, General specification table)
.
Note:
The power must be supplied by a fused or overload-protected power supply.
3.
Connect the power supply to the instrument.
4.
Switch the power supply on.
5.
Check that the front panel display shows the power-up sequence
(Ref: section 3.2,
Power-up sequence)
.
Summary of Contents for PACE5000
Page 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Page 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Page 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Page 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...