Z Series RO System by GE Osmonics
5.14 RO Fails To Draw Disinfectant
1.
Disinfect Inlet Valve Closed
a. Open
valve.
2.
Housing On Submersible Pump Contains Air
a. Purge air, see cleaning or disinfect procedure.
3.
Disinfect Line Kinked Or Plugged
a. Clear disinfect line.
4.
Air Leak On Disinfect Draw Line
a. Tighten
fittings.
b. Check O-ring in Disinfect Inlet Valve.
5.
Motor Fails To Turn On
a. Disinfect key switch inoperative.
b. Remote monitor (if equipped) key switch not turned.
c. D.A.M. (if equipped) timer not timed out.
d. See 5.1.
6.
Solenoid Failure In Open Position (Fluid Level In Tank Will Increase)
a. See
5.4.
5.15 Product Divert Controller and Valve Malfunction
1.
Connections
a. Ensure the Product Divert Controller power cord is connected to the appropriate
outlet.
b. Check outlet and circuit breaker/fuse panel for power.
c. Verify the wire lead connections inside the Product Divert Controller, and to the
Divert Valve and RO controller are correctly and securely attached. Reference the
wire connection diagrams in Chapter 6 for a visual description of wire connection
locations. Restore connections as required.
d. Check fuse and relay, replace as necessary.
2.
Divert Valve
b. Verify correct valve movement.
c. Verify the divert valve is operational using separate power source. Replace valve
as necessary.
3.
Indicator
a. Verify the indicator lamp has not burnt out with a continuity check across the
terminal connections. Replace indicator as necessary.
1238339b – 16Oct03
5 - 11
Troubleshooting
Summary of Contents for Osmonics Z-10000
Page 2: ......
Page 6: ...This page intentionally left blank 1238339b D D 16 Oct 2003 ...
Page 27: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 1 21 General Information ...
Page 43: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 3 5 System Operation ...
Page 72: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 4 28 Routine Maintenance ...
Page 84: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 5 12 Troubleshooting ...
Page 107: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 23 Drawings Z Series RO Fluid Panel ...
Page 119: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 6 35 Drawings ...
Page 126: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 7 6 Spare Parts List ...
Page 127: ......