Z Series RO System by GE Osmonics
4.4.1.1 Formaldehyde Disinfection Procedure
NOTE:
If your RO system is equipped with a "submersible" high-pressure pump, the
pump/motor housing must be purged of all air during the start-up procedure. Turn on the
RO system and open the small valve on the top end of the pump/motor housing very
slowly to purge. Air should be heard escaping to the atmosphere. Close the valve when
water is seen flowing from this valve.
1.
Turn off the RO unit power switch.
2.
Label machine with appropriate warning signs such as “Do Not Use / Contains
Disinfectant” to prevent use of the RO until it is properly rinsed. Warning signs
should be placed at each use point if the loop piping is being disinfected.
3.
Remove the RO product line from the Product Divert Valve Assembly (RO
Product Water connection) and connect to one of the upper connection ports on
the disinfect tank.
4.
Prepare a 4% solution of formaldehyde by mixing 1 gallon of 37% formaldehyde
to 9 gallons of RO product water.
a. Use the RO unit to fill the disinfect tank with 9 gallons of product water and
turn off the RO.
b. Add 1 gallon of 37% formaldehyde to the disinfect tank and mix well.
NOTE:
Ten gallons of solution is adequate for the V-Series RO; if the unit is directly feeding a
distribution loop and the loop is to be disinfected at the same time, a larger volume of
solution may be required.
5.
Connect a disinfect line hose between the Disinfect Inlet Valve on the RO and the
lower connection port on the disinfect tank.
6.
Remove the RO waste line from drain and connect to an upper connection port on
the disinfect tank.
7.
Open the Disinfect Inlet Valve and the Waste Flow Valve (counter clockwise) on
the RO unit. Direct Feed RO units will also require the loop return ball valve,
mounted vertically just prior to pump housing, to be closed during disinfection.
8.
Loop piping disinfection will require the product line to be removed from the
disinfect tank and connected to the loop before proceeding. Direct Feed RO units
having the loop disinfected will require the loop return ball valve, mounted
vertically just prior to pump housing, to be open during disinfection.
1238339b – 16Oct03
4 - 17
Routine Maintenance
Summary of Contents for Osmonics Z-10000
Page 2: ......
Page 6: ...This page intentionally left blank 1238339b D D 16 Oct 2003 ...
Page 27: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 1 21 General Information ...
Page 43: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 3 5 System Operation ...
Page 72: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 4 28 Routine Maintenance ...
Page 84: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 5 12 Troubleshooting ...
Page 107: ...Z Series RO System by GE Osmonics 1238339b 16Oct03 6 23 Drawings Z Series RO Fluid Panel ...
Page 119: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 6 35 Drawings ...
Page 126: ...Z Series RO System by GE Osmonics NOTES 1238339b 16Oct03 7 6 Spare Parts List ...
Page 127: ......