Section 3.3 Heat output
3.2.1 Atmospheric pressure
Ensure that the in-use and storage locations meet the atmospheric pressure minimums and
maximums listed in
Ensure that the in-use and storage location conditions do not fluctuate more than the maximum
rates of change given in
Section 3.3 Heat output
Below are the maximum power and thermal values in summary form:
For reference, a tightly closed room of 3 m x 3 m (10 ft x 10 ft) with a 2.4 m (8 ft) ceiling would require
2100 BTU/hour to raise the temperature by 12.2°C (10°F).
Section 3.4 Audible noise
Noise limits assume that all system covers are installed, and assume background noise of 40 dBA.
Measurements are taken 1 m (3.3 ft) from the equipment. The in-use measurement is taken when
the unit is being driven. Tube rotor and exposure tone noise are not included in these
measurements.
Maximum operating noise generated by the equipment is listed in
Table 3-4 Atmospheric pressure requirements
In-use
Storage
Min.
Max.
Min.
Max.
70.1 kPa
106 kPa
70.1 kPa
106 kPa
Table 3-5 Atmospheric pressure change rate limits
In-use
Storage
< 1.8 kPa/hour
< 76 kPa/hour
Table 3-6 Optima XR200amx/XR220amx heat output
State
Watts
BTU/hour
Idle
143
488
Drive
275
938
Charge
261
890
Expose
474
1616
Expose
611
2084
Table 3-7 Audible noise produced by the system
In-use
Idle
Upper limit: 55 dBA
Upper limit: 50 dBA