Section 1.5 Product storage and handling requirements

1.4.4  Minimum elevator depth requirements

See 

1.4.1 Product dimensions and weights

 an

1.4.2 Shipping dimensions and weights

.

Section 1.5  Product storage and handling requirements

Refer to 

Chapter 3 Environmental (HVAC) requirements

 for temperature, humidity, and 

atmospheric pressure requirements.
The maximum incline that the system can navigate is 7 degrees, and the maximum threshold height 
over which the system can drive is 2.5 cm (1 in).

The system is not designed to travel over rough or bumpy ground. Take care when moving 
the system to avoid sharp bumps, or steps larger than the 2.5 cm (1 in) threshold limit. 

Section 1.6  Pre-installation results

After the pre-installation work is completed, you should be able to check off each of these items:

1.

Radiation physicist consulted about radiation protection

2.

Electrical service in place at the ratings specified

3.

Charging/downloading area size adequate for equipment

4.

Secure location to store loose items identified

5.

Network requirements met or exceeded

6.

If wireless option is purchased, wireless network capabilities met

7.

Network/workflow information collected

8.

IP addresses for DICOM and broadband identified

9.

Broadband access provided for InSite ExC connection

10. If tablet holder option is purchased, the tablet is ready to install.                                  

Summary of Contents for Optima XR200amx

Page 1: ...GE Healthcare Optima XR200amx XR220amx Pre Installation Manual Direction 5336112 1EN Revision 3 Front cover...

Page 2: ...ned in Annals Number 26 of the ICRP and with applicable national standards ATENCION LOS APARATOS DE RAYOS X SON PELIGROSOS PARA EL PACIENTE Y EL MANIPULADOR CUANDO LAS NORMAS DE PROTECCION NO ESTAN OB...

Page 3: ...jeli ovaj servisni priru nik zanemarite li ovo upozorenje mo e do i do ozljede davatelja usluge operatera ili pacijenta uslijed strujnog udara mehani kih ili drugih rizika V STRAHA CS Tento provozn n...

Page 4: ...e est disponible uniquement en anglais Si le technicien d un client a besoin de ce manuel dans une langue autre que l anglais il incombe au client de le faire traduire Ne pas tenter d intervenir sur l...

Page 5: ...ens serviceleverand r har bruk for et annet spr k er det kundens ansvar s rge for oversettelse Ikke fors k reparere utstyret uten at denne serviceh ndboken er lest og forst tt Manglende hensyn til den...

Page 6: ...l s Si el encargado de mantenimiento de un cliente necesita un idioma que no sea el ingl s el cliente deber encargarse de la traducci n del manual No se deber dar servicio t cnico al equipo sin haber...

Page 7: ...arks of GE Healthcare WPA and WPA2 are trademarks of the Wi Fi Alliance All other products and their name brands are trademarks of their respective holders COPYRIGHTS All material copyright 2011 by Ge...

Page 8: ...e its own specially trained field engineers All of GE s electrical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only...

Page 9: ...E Healthcare sales or service representatives GE personnel please use the GE Healthcare complaint handling process to report all omissions errors and defects in this publication DANGER Risk of Explosi...

Page 10: ...XR200amx under Shipped Separately Refer to ECO 2162103 Chapter 1 Section 1 3 2 added customer s responsibility for tablet holder option Chapter 1 Section 1 6 added a item of tablet holder in the chec...

Page 11: ...nal word The safety alert symbol is the triangle with an exclamation mark within it It Is always used next to the signal word to indicate the severity of the hazard Together they are used to indicate...

Page 12: ...cons symbols may be used to make you aware of specific types of hazards that could possibly cause harm keep_up fragile static_elec keep_dry general torque ce magnetic impact heat pinch explosive crush...

Page 13: ...ples of information possibly stated previously Comment Represents additional information that may or may not be relevant to your situation Page Layout Headers and footers in this publication are desig...

Page 14: ...Example Fixed Input This paragraph s font represents fixed input It is computer input that is typed in via the keyboard Typed input that does not vary from application to application or system to sys...

Page 15: ...hipments 23 1 4 3 Door size requirements 23 1 4 3 1 System in crate 23 1 4 3 2 System in normal use not in crate 23 1 4 4 Minimum elevator depth requirements 24 Section 1 5 Product storage and handlin...

Page 16: ...e 16 Table of Contents Section 5 2 Wireless networks 31 5 2 1 Access points and network capabilities 31 Chapter 6 Network workflow information 33 Section 6 1 Network workflow information 33 6 1 1 Faci...

Page 17: ...anual to determine whether the site is ready for the installation to begin This is an important process that helps verify site readiness 1 1 1 Pre installation tasks Pre installation is a cooperative...

Page 18: ...r Optima XR220amx only Service keyboard Software CD External CD DVD ROM drive Operating System firmware USB drive Tablet holder assembly optional Optima XR220amx and Optima XR200amx Shipped separately...

Page 19: ...lity includes providing 1 A clean and safe work environment for installation of the product finished floor ceiling walls and proper lighting 2 A secure location for storing loose items such as the ser...

Page 20: ...20amx 119 3 cm 47 in 55 2 cm 21 7 in 194 0 cm 76 4 in standard column or 178 8 cm 70 4 in short column 404 kg 891 lb Figure 1 2 FlashPadTM detector Optima XR220amx or Optima XR200amx with digital upgr...

Page 21: ...GE HEALTHCARE OPTIMA XR200AMX XR220AMX DIRECTION 5336112 1EN REVISION 3 PRE INSTALLATION MANUAL Chapter 1 General requirements Page 21 Figure 1 2 System dimensions Figure 1 3 FlashPadTM detector 552...

Page 22: ...pping dimensions when reviewing door size hallway width corner angles and elevator size Components not listed are of negligible size and weight 1 4 2 1 United States shipments Component Length Width H...

Page 23: ...the system during normal use the minimum door width is 61 cm 24 in and the minimum door height is 193 cm 76 in See the room diagram in Figure 1 6 Figure 1 6 System room diagram Component Length Width...

Page 24: ...to travel over rough or bumpy ground Take care when moving the system to avoid sharp bumps or steps larger than the 2 5 cm 1 in threshold limit Section 1 6 Pre installation results After the pre insta...

Page 25: ...ial construction requirements Section 2 1 Radiation protection Because X ray equipment produces radiation special precautions may need to be taken or special site modifications may be required GE Heal...

Page 26: ...GE HEALTHCARE OPTIMA XR200AMX XR220AMX DIRECTION 5336112 1EN REVISION 3 PRE INSTALLATION MANUAL Page 26 Section 2 1 Radiation protection...

Page 27: ...ble 3 2 Section 3 2 Altitude Storage values refer only to equipment that is still in shipping containers If the equipment is partially or completely installed refer to in use values Ensure that the in...

Page 28: ...C 10 F Section 3 4 Audible noise Noise limits assume that all system covers are installed and assume background noise of 40 dBA Measurements are taken 1 m 3 3 ft from the equipment The in use measure...

Page 29: ...t of non conformity may be the responsibility of the institution and or their contractor Section 4 1 Power requirements 4 1 1 Electrical requirements Customer supplied electrical outlet United States...

Page 30: ...is drawn from the batteries even if the system is connected to wall power Power consumption during battery charging assuming the system is not being used during charging is as follows Draw is less tha...

Page 31: ...onnection must be 802 3 compliant 10 100 1000 Base T Ethernet and there must be at least one RJ45 jack dedicated for Static IP to connect the new equipment on the LAN Section 5 2 Wireless networks If...

Page 32: ...CHAPv2 EAP FAST EAP SIM PEAP GTC Encryption WPA2TM PSK AES Authentication EAP TLS Frequency band and operating channels If the installation country is not listed check with the local regulatory agency...

Page 33: ...________________ Wired network contact ________________________________ Phone __________________ Wireless network contact ______________________________ Phone __________________ DICOM device contact _...

Page 34: ...____________________ What is the password _____________________ What is CA certificate if required _____________________ What is Client certificate if required _____________________ What is Private ke...

Page 35: ...conflict If this conflict occurs contact a GE Healthcare Service Representative to change the internal IP addresses used by the system The host name is the network s name for the Optima XR200amx XR22...

Page 36: ...ach patient Image handling Optima XR220amx only What percent of images acquired are reviewed via soft copy ______ What percent of images acquired are printed ______ After the diagnostic images are acq...

Page 37: ...tware firmware version _________________________ Network IP address _ _ _ _ _ _ _ _ _ _ _ _ The host name is the name that appears on the screen that users will use to identify this device Host name _...

Page 38: ...PPS server receives the messages sent by the Optima XR200amx or Optima XR200amx with digital upgrade system These messages consist of information such as when the exam started and closed how many imag...

Page 39: ...ay retrieve studies series images For patient the user may retrieve all of the study attributes plus a patient s entire image collection Query retrieve by Study Patient A storage commitment provider c...

Page 40: ...erver Yes No If yes provide AE title __________________________ Port number ___________ Being a query retrieve service class provider allows the Optima XR220amx or Optima XR200amx with digital upgrade...

Page 41: ...services that remote host devices may offer Each of the services will have its own AE application entity title and port number The AE title is the name given to a service or application provided by a...

Page 42: ...Yes No Indicate the highest level of DICOM conformance for this device If the device does have some level of DICOM conformance return a copy of the DICOM Conformance Statement with this completed form...

Page 43: ...ance for this device If the device does have some level of DICOM conformance return a copy of the DICOM Conformance Statement with this completed form DICOM compatibility does not guarantee that all f...

Page 44: ...GE HEALTHCARE OPTIMA XR200AMX XR220AMX DIRECTION 5336112 1EN REVISION 3 PRE INSTALLATION MANUAL Page 44 Section 6 1 Network workflow information...

Page 45: ......

Page 46: ...46 2011 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 3000 N Grandview Boulevard Waukesha Wisconsin 53188 USA www gehealthcare com...

Reviews: