Revision D
MAC
1200 resting ECG analysis system
5-3
2012250-095
Troubleshooting: Basic Information
Basic Information
Repairs may only be conducted by approved service personnel.
For all the following points, see Chapter 4,
that need to be observed.
NOTE
The master password used to override all other passwords is SYSTEM.
Device cannot be switched on even though power plug is plugged in
The green power lamp LED is off and device cannot be switched on.
Is the power cable defective or not plugged in correctly?
Are the primary fuses in the mains input module defective?
Is the connector from the mains input module to the mains switching power
supply plugged in correctly or cable defective?
Is the connector from the mains switching power supply to connector
POSUP/
on the PCB Control plugged in correctly or cable defective?
Is the keypad via connector
KEYB/
on PCB Control plugged in correctly?
It there a 26.5 V on the connector
POSUP
/?
If no: mains switching power supply defective.
If yes: PCB control defective or keypad defective.
The green power lamp LED is on, but device still cannot be switched on.
Is the PCB control defective?
Is the keypad defective?
Device cannot be switched on when running on battery power
Is the battery depleted?
Plug in power plug. The green power lamp LED should glow and the
device can be activated.
“Device cannot be switched on even though power plug is
By connecting the power plug, rapid charging is activated (as indicated
with LED1 on the PCB battery charge).
If not: PCB battery charge is defective.
With power plug plugged in, let the unit charge for 10 minutes, then
disconnect power plug. Can the device be switched on and can a recording
be started?
If yes: Function OK. Continue loading with plugged mains plug.
If not: Is the battery disconnected or cable defective?
Is the battery defective (no capacity)?
Is the PCB battery charge defective?
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......