Repair Instructions: Replacing Components
Replacing Components
For all the following points, the
on page 4-3 need to be
observed.
WARNING
The device has to be switched off.
Opening the Device
To open the device, follow this sequence.
1.
Release the five fastening screws on the underside of the device.
2.
Open the paper feed flap.
3.
Carefully raise the upper shell of the housing.
4.
Disconnect the keypad cable by opening the connector KEYB/ on the PCB
Control.
The display remains in the lower shell of the casing.
During reassembly, ensure that no cables are pinched.
Replacing the Primary Fuses in the Mains Input Module
WARNING
Only use fuses indicated on the rating label. (See Chapter 8,
for details.)
1.
Switch off unit and disconnect mains plug.
2.
Insert small flat blade screwdriver sideways into the lower section of the power
connector.
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......