Revision D
MAC
1200 resting ECG analysis system
7-3
2012250-095
Service and Maintenance: Technical Inspections/Preventative Maintenance
Technical Inspections/Preventative Maintenance
It is recommended that a technical inspection/preventative maintenance check is
performed once a year. The following checks/tests are to be performed.
Check device and accessories for mechanical defects which impair their
function.
Perform a function check as detailed in Chapter 3,
Ensure that labels and inscriptions on the device relating to safety are clearly
legible.
Measure the Protective Earth Resistance (see
Measure the enclosure leakage current (see
“Enclosure Leakage Current Test”
Measure the patient leakage current (see
“Patient Leakage Current Test”
page 7-9).
WARNING
The following checks may only be performed by persons whose
training, knowledge and practical experience enable them to carry
out such checks reliably and correctly.
The operational and functional reliability of the device is checked using the
information in this chapter, which serves the experienced technician when checking
the device.
A knowledge of device operation as detailed in the “Operator's Manual” is assumed.
The checks and tests are based on the testing instruments given in this section.
The tests should be carried out using the customer's accessories, so that defective
accessories are also detected automatically. If other testing instruments are used
besides those mentioned, the items on the checklist and tolerance specifications may
need to be modified.
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......