Revision D
MAC
1200 resting ECG analysis system
2-7
2012250-095
Functional Description: PCB Control CS_CI
PCB Control CS_CI
The PCB Control CS_CI is the main board of the MAC
1200 resting ECG analysis
system. The main board accommodates the control functions of the device, except
the line power supply and the NC battery-charging controller.
Generation of Internal Power Supplies
Logic Power 5 V and +3.3 V
The processor core and the memory is a fully +3.3 V design. The interface functions
to display, to RS-232, to the printing unit and to the ECG acquisition unit are
supplied with +5 V.
Both the +3.3 V and the +5 V are generated separately by a clock-rated voltage
controller. By using the “adjustable version” of the voltage controller, both supplies
are based on the same voltage controller. The output voltage is determined by
appropriate dimensioning of the voltage divider at the feedback input of the voltage
controller.
Standby 5 VSTB and +3.3 VSTB
The standby 5VSTB supplies the ON/OFF Control, the + 3.3 VSTB
buffers the patient ECG memory when the device is turned off. Both supplies are
generated together from one low-power linear regulator with low quiescent supply
current.
Analog 12 V
+12 V is used for the heating control and temperature monitoring of the printing
unit, which requires only a current of a few milliamperes. This supply is generated
by a linear regulator with feedback input.
FPGA Power 2.5V +1.25V
+2.5 V is the auxilliary source of power and +1.25 V is the main power supply for
the FPGA internal logic. These voltages are generated by Dual Output Low Dropout
Regulator.
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......