7-10
MAC
1200 resting ECG analysis system
Revision D
2012250-095
Service and Maintenance: Technical Inspections/Preventative Maintenance
Test has failed if the measured values are greater than 50 µA.
Electrical Diagram for Patient Leakage Current Test
For the protection of the test person, the following values of resistor R may be used.
Type BF
22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
Type CF
100 kOhm (220 to 240 V)
022A
Safety Tester
[M]
GND
Power Cord
Patient
Cable or
Test Plug
NORM
High
GND
Low
Un
it Un
der T
e
st
Power Cord
R
RVS
Summary of Contents for MAC 1200
Page 10: ...vi MAC 1200 resting ECG analysis system Revision D 2012250 095 ...
Page 11: ...Revision D MAC 1200 resting ECG analysis system 1 1 2012250 095 1 The Basics ...
Page 12: ...1 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 19: ...Revision D MAC 1200 resting ECG analysis system 2 1 2012250 095 2 Functional Description ...
Page 20: ...2 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 45: ...Revision D MAC 1200 resting ECG analysis system 3 1 2012250 095 3 System Test Function ...
Page 46: ...3 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 69: ...Revision D MAC 1200 resting ECG analysis system 4 1 2012250 095 4 Repair Instructions ...
Page 70: ...4 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 79: ...Revision D MAC 1200 resting ECG analysis system 5 1 2012250 095 5 Troubleshooting ...
Page 80: ...5 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 85: ...Revision D MAC 1200 resting ECG analysis system 6 1 2012250 095 6 Adjustment Instructions ...
Page 86: ...6 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 89: ...Revision D MAC 1200 resting ECG analysis system 7 1 2012250 095 7 Service and Maintenance ...
Page 90: ...7 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 101: ...Revision D MAC 1200 resting ECG analysis system 8 1 2012250 095 8 Parts List ...
Page 102: ...8 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 111: ...Revision D MAC 1200 resting ECG analysis system 9 1 2012250 095 9 Specifications ...
Page 112: ...9 2 MAC 1200 resting ECG analysis system Revision D 2012250 095 For your notes ...
Page 123: ...Revision D MAC 1200 resting ECG analysis system 10 1 2012250 095 10 Drawings ...
Page 150: ......