GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Chapter 10 -
Page 10-19
10-6-5-4
Data Sheet for Chassis Source Leakage Current
The test passes when all readings measure less than the value shown in
. Record all data
on the PM Inspection Certificate.
Table 10-18 Typical Data Sheet for Chassis Source Leakage Current
Unit Power
Tester Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1
Probe
Connector
Test 2
caster
Test 3
CRT
Optional
Test 4
Optional
Test 5
Enter Name of tested peripheral here:
ON
NORM
OPEN
ON
NORM
CLOSED
ON
REV
OPEN
ON
REV
CLOSED
OFF
NORM
OPEN
OFF
NORM
CLOSED
OFF
REV
OPEN
OFF
REV
CLOSED