LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-36
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
7.) ARM FOLD COVER
Unscrew 1 screw (2329677, TAP M4X16) to remove the Arm fold cover.
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-5-4
Mounting procedure
Install the each of the new part in the reverse order of removal procedure.
Figure 8-69 Removing the Arm fold cover
Table 8-6
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-5. Equipment
passed all required tests and is ready for use. “