GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 7-22
Section 7-1 - Overview
7-7-4-3
PI128 II (LOGIQ™ P6/P6 Pro)
These programs are provided for testing the PI128 II assy.
NOTE:
A Convex Probe is needed for PI128 II Channel Test.
7-7-4-4
CWD64 (LOGIQ™ P6/P6 Pro)
These programs are provided for testing the CWD64 assy.
Table 7-4
FRU Test, PI128 II Diagnostics Menu
Menu
Descriptions
User
Level
Tools
Full Test
Performs all of the PI128II assy tests listed below
All
None
Assy Revision
Test
Test EEPROM.
Reads the assy revision of the PI128 II assy.
All
None
FEBUS Test
Checks if the DSP in SYSCONML can access the PI128 II assy via FEBUS.
All
None
SHV Test
Send the command to SHV resister in PI128 II assy to read SHV status.
All
None
OPEN Test
Checks whether probe connected or not on PI128 II assy.
All
Probe
PROBE CODE
Checks if the 8 bits of the Probe Code in the probe connector is properly connected.
If connected probe is available, display code and Probe name.
All
Table 7-5
FRU Test, CWD64 Diagnostics Menu
Menu
Descriptions
User
Level
Tools
Full Test
Performs all of the CWD64 board test listed below
All
None
Assy Revision
TEst
Test EEPROM.Send test pattern to write to EEPROM and readDisplays the board
revision of the CWD64 assy.
All
None
USC Interface
Test
Checks if the DSP in SYSCONML can access the CWD64 assy via FEBUS.
All
None
FEBUS Test
Allows SYSCONML to access FEBUS interface in the CWD64 in order to check if FE
bus line is normal.
All
IQ Data Read
Send the command to perform CW scanning, read IQ data from assigned memory,
analyze the data with FFT algorithm.
All
CWD TPG Test
Read write the TPG Register in CWD64 via FEBUS.
Probe
Connection
Checks whether Dedicated CWD probe connected or not.
All
Channel Test
Generate test pattern, perform 48 channel scanning in CWD64, read the data and
analyze it.
All