GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 4-16
Section 4-3 - General Procedure
4-3-11
Backup and Restore Database, Preset Configurations and Images
NOTE:
When upgrading from R1.0.x to R2.0.x or R3.0.x , preset cannot be transferred. Refer to the
table below.
4-3-11-1
Formatting Media
1.) To format the backup media, CD-R or DVD-R, select the UTILITY key on the Front Panel.
2.) Select CONNECTIVITY, then RERNOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-13. It is best to use all capital letters with no
spaces or punctuation marks.
Table 4-4
LP6/LP6pro BT11(R3.x.x) compatibility
Preset
Compatibility
R1.x.x
R2.x.x
R3.x.x
Patient Archive
Yes
Yes
Yes
Report Archive
Yes
Yes
Yes
Imaging Presets
No
No
No
Connectivity Configuration
No
No
No
Measurement Configuration
No
No
No
Comment/Body Pattern Libraries
No
No
No
Report Templates (Same Software Version Only)
No
No
No
3D/4D
No
No
No
Key Macro
No
No
No
All Others
No
No
No