Parameter
Suggested Value or Source of Data
1
Default
Allowable range
2
PD average level trend alarm
0 (disabled)
0 to 65000
PD count level warning
0 (disabled)
0 to 65000
PD count level alarm
0 (disabled)
0 to 65000
PD count trend warning
0 (disabled)
0 to 65000
PD count trend alarm
0 (disabled)
0 to 65000
PDI level warning
1000 mW
0 to 65000
PDI level alarm
2000 mW
0 to 65000
PDI trend warning
0 (disabled)
0 to 65000
PDI trend alarm
0 (disabled)
0 to 65000
Table 4: Partial Discharge Alarm Settings