Patient Temperature Control Mode
In baby mode, the system performs the following adjustment every 30 seconds. The Baby Condition
column indicates the difference between the baby skin temperature and the baby mode set
temperature.
Baby Condition
Heater Output Adjustment
Warm by 0.15°C or more
Decrease by 5%
Warm by 0.06°C to 0.14°C
Decrease by 1%
Warm by 0.05°C or less
No adjustment
Cold by 0.05°C or less
No adjustment
Cold by 0.06°C to 0.14°C
Increase by 1%
Cold by 0.15°C to 0.24°C
Increase by 5%
Cold by 0.25°C to 0.34°C
Increase by 10%
Cold by 0.35°C or more
Increase by 15%
Specifications
The product is designed to meet a life span of 8 years. However, with proper maintenance and
repairs, the service life can be extended as long as service parts are available.
The maximum patient weight is 40 kg (88 lbs).
During the seven minute warmup period, the radiant heater is at 100% power (~27 mW/cm
2
). After
the radiant heater has been running in warmup mode for seven minutes, the heater power setting
drops to 50% (10 to 15 mW/cm
2
).