Panda iRes Warmer
M1110737 006
© 2007 General Electric Company, all rights reserved
WARNING:
Only use Ohmeda probes; other manufacturers probes are not calibrated to GE Healthcare
equipment. Using probes from other manufacturers may cause inaccurate temperature
readings, may not comply with safety standards, and will void your GE Healthcare
equipment warranty.
WARNING:
When using phototherapy lamps
, the probe must be directly in the path of the radiant heat
of the lamp; do
not
place the probe in an area shielded from the lamp’s light. The photo
-
therapy lamp may raise the baby’s skin temperature
.
WARNING:
Do
not
use rectal temperatures to control the baby’s temperature
.
WARNING:
Do
not
remove the probe from its storage bag until required for use. Replace the probe if
the cable or tip becomes damaged.
WARNING:
Do
not
re-use a disposable probe. Cleaning and reusing a single use probe may damage
the probe and result in inaccurate readings.
Attaching the Skin Temperature (continued next page)