1-15
Panda iRes Warmer
M1110737 006
© 2007 General Electric Company, all rights reserved
Alarms
Illustration
Description
1
2
3
4
There are two types alarms:
Low Priority Alarms
are indicated by a blinking red
alarm light, a red alarm message on the central
alarm area of the display, and an intermittent
audio tone.
High Priority Alarms
have the same blinking red
alarm light and red alarm message. The difference
is that the audio tone for high-priority alarms is
continuous.
Alarm features
Feature
number
Description
1
Alarm light
2
Motion sensors
3
Alarm silence key
4
Alarm message
To silence the alarm.
There are two methods of
silencing the alarm:
1. Press the yellow button at the top of the control
panel.
2. Use the Hands Free Alarm Silence (see below).
When you silence an alarm, the text message gets
smaller and moves to the top of the screen and the
red alarm light goes from blinking to solid. When
the alarms have been silenced, the audio tone is
temporarily turned off, but the warmer remains
in alarm status until the alarm condition clears.
When the silence period expires, the audio tone
will resume, but the silence period varies with the
alarm (for greater detail about alarm priorities and
silence periods, see the Alarm Table at the end of
this chapter). If there are multiple alarms, the text
messages alternate every 2 seconds.