FineLINE 35
oligo
column
Operating Instructions
28-9649-57
AA
31
Maintenance 5
Component maintenance 5.5
Bottom filter unit replacement
Adapter filter unit replacement
Step
Action
1
Use the wrench supplied to remove
the old bottom filter unit.
2
Wet the new bottom filter unit with ethanol and insert into the column tube.
3
Tighten the bottom filter.
Step
Action
1
Unscrew the two screws that hold the
adapter filter unit and remove it from
the adapter rod.
2
Gently pull out the filter unit.
Summary of Contents for FineLINE 35 oligo
Page 1: ...GE Healthcare Operating Instructions Original instructions FineLINE 35 oligo column ...
Page 2: ......
Page 4: ...Table of Contents 4 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Page 18: ...3 Installation 18 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Page 34: ...6 Troubleshooting 34 FineLINE 35 oligo column Operating Instructions 28 9649 57 AA ...
Page 43: ......