Operating Instructions
28-9649-57
AA
29
Maintenance 5
Disassembly and assembly 5.4
Assembly
Step
Action
1
Wet the bottom filter with ethanol.
2
Insert the bottom filter unit into the
column tube and tighten it with the
supplied wrench (a).
3
Wet the adapter filter unit and the adapter rod with ethanol.
4
Move the adapter lid (b) to its lower stop position.
5
Insert the adapter into the column tube so that the slot in the adapter lid
is positioned directly above “1” on the column top end-piece (c).
6
To lock the adapter,
turn the adapter lid
clockwise so that the
slot is positioned
directly above “2” on
the top end-piece (d).