1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
. This example shows one function:
•
Thermocouple (TC) is set to source a temperature. The
Utility
is set to
Switch Test
. The
Automation
is set to
Ramp;
see
Section 2.11.2
.
2.
Complete the electrical connections. Because it is a
Channel 1 function (TC), you must have the switch
connections on Channel 2 (CH2).
3.
For the
Ramp
process, set “High” and “Low” values that are
applicable to the switch value. Then, to get an accurate
switch value, set a long “Travel” period.
Use Start/Stop to start and stop the “Ramp” cycle. If
necessary, supply the output values in the opposite direction
until the switch changes condition again.
4.
The display shows:
•
the values to open and close the switch
•
the hysteresis value
5.
To do the test again, use the reset button.
3.3 Measure
pressure: IDOS option
Optional item.
An IDOS Universal Pressure Module (UPM)
uses Intelligent Digital Output Sensor (IDOS) technology to
measure the applied pressure and supply the data to an IDOS
instrument.
Before you use an IDOS module, read the user manual: K0378,
Druck IDOS UPM.
Note:
To attach an IDOS module to the DPI 620 calibrator, you
must have an adaptor: IO620-IDOS-USB.
1
3
2
4
Start/Stop
Closed
Open
Reset
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......