[EN] English - K0449
Prepare the instrument 2-9
Issue 1
2.10.3 Procedures to make Task selections
When you first use the calibrator, there are default
measure
and source
functions on the display:
•
CH1 settings: RTD source, RTD type is PT100, scale is °C;
Automation is
Nudge
; see
Chapter 3
.
•
CH2 settings: Current measure, see
Chapter 3
.
Procedure Overview
Use the
Task
menu to complete these procedures:
•
Set the calibrator functions you want to use on the display:
item ;
see
Section 2.10.4
.
a. CH1: Channel 1 electrical function: You can choose one
option:
b. CH2: Channel 2 electrical function: You can choose one
option:
c. * Pressure function (P1): for the PV 62x pressure
stations, refer to user manual - K0457; for the MC 620
module carrier, see
Chapter 4
.
d. * Pressure function (P2): the MC 620 module carrier can
use P1 and/or P2; see
Chapter 4
.
e. * IDOS function: Universal pressure module; see
Chapter 3
.
f. HART function: HART device communications; see
Chapter 8
•
If applicable, change the
Units
for the function: item
;
see
Section 2.10.5
CH1:
source
CH2:
measure
TAP
2
1
C
B
A
None
Resistance
Current
RTD
Voltage
CJ
Millivolts
AC Volts
TC (°C/°F)
AC mV
TC (mV)
AC Probe
Frequency
Observed*
None
Voltage
Current
Millivolts
Current (24V)
Observed*
A
Note:
* Observed (Only
available as a source
option). Use this option to
make a manual record of
the readings on another
instrument; see
Section 2.11.3
B
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......