Issue 1
3-2 Electrical and IDOS operations
K0449 - [EN] English
•
If necessary, change the
Process
for the CH1 and/or CH2
measure functions: item
This includes:
Tare
,
Alarm
,
Filter
,
Flow
,
Scaling
; see
Section 2.11.1
.
There are more optional
Settings
for the
TC
,
Frequency
, and
RTD
functions.
•
If necessary, change the
Automation
options for the CH1
and/or CH2 source functions: item
This includes:
Nudge
,
Span Check
,
Percent Step
,
Defined Step
,
Ramp
; see
Section 2.11.2
There are more optional
Source Settings
for the
TC
,
Frequency
,
and RTD
functions.
There are other
Settings
for
Observed
functions; see
Section 2.11.3
.
•
When all the software selections are complete, make the
applicable electrical connections.
3.2.2 Example procedure: Measure or source current
These examples (A and B) show Channel 1 (CH1) set up to
measure or source a current with external loop power.
Note:
If you use the Channel 2 (CH2) connectors, you can use
Channel 2 to measure or source these ranges with internal or
external loop power (internal loop power = 24 V).
1
D
TAP
D
1
E
TAP
E
2
Maximum: 30V
Source current on
channel 1 (CH1)
Range: 0 to 24 mA
Automation
:
Nudge
(
Section 2.11.2
)
2
Maximum: 30V
1
B
3
Measure current on
channel 1 (CH1)
Range: ±55 mA
A
1
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......