Issue 1
8-6 HART
®
device operations
K0449 - [EN] English
2.
Set the
HART Resistor
in the
Advanced
menu:
3.
Complete the electrical connections and continue with the
HART menu operation; see
Section 8.5
.
8.4.3 Communicator attached to a network
In this example, the calibrator connects directly to a network.
There must be a 250
:
resistor in series with the loop power
supply and the HART device.
1.
Set the function for HART:
e
2
Set to On
Section 2.9
a
b
c
d
Home display
TAP
No functions
on channel 2 (CH2)
HART Function is set
250
:
resistor is set to
Off
d
1
a
b
c
Set HART
Section 2.10.4
Home display
TAP