6
To Start
K0394 Issue 3 - [EN] English
Page 5
Menu Description
Steps
Result/Subsequent steps
1
2
C _ _ _ _
=
Calibration:
To continue, set
the correct calibration
access code = last four digits
of S/N. *******
C0
(correct the zero offset value)
➤
C2
(do a two-point pressure calibration)
➤
V2
(do a two-point voltage calibration), refer to
menu operations.
A OFF
= Set low/high alarm:
Set to On or OFF.
On
➤
000.0
↓
➤
100.0
↑
Set a value for the low and/or high alarm (0 to
105% FSO).
OFF
=
Supply voltage output
(Vout):
Set to OFF, P-V, or US.
P-V:
Vout is proportional to the pressure value on
the display. Make sure the Vout scale factor is
correct.
US
➤
000.0: Set a Vout value (0 to 100%) to control
an external pressure regulator. Make sure the
Vout scale factor is correct.
S 1.00
=
Set Vout scale factor:
A
Vout adjustment.
If applicable, set a new Vout scale factor (0.01 to
9.99), factory value = 1.00.
Au On
=
Set automatic power OFF:
Set to On or OFF.
On
➤
Au 15 : Set the period for automatic power
OFF (1 to 99 minutes), factory value = 15 minutes.
L OFF
=
Set lock code:
A menu
protection facility. Set to On
or OFF.
On
➤
L 000 : Set a new lock code (if necessary),
factory code = 000.
Sc 02
=
Set scan rate:
A rate that
the DPI 104 uses to take
pressure samples.
Set an applicable rate (02 to 10 Hz),
factory value = 02 Hz.
OFF
=
Monitor external IDOS:
Set
to On or OFF.
Monitor function is set on or off.
FS
↓
=
Set FSO low register:
To set
a different range for these
functions: analog display, %,
alarm.
Set a value for the low end of the range (refer to
Table 8, page 16).
Factory value = Factory calibration value.
FS
↑
=
Set FSO high register:
To set
a different range for these
functions: analog display, %,
alarm.
Set a value for the high end of the range (refer to
Table 8, page 16).
Factory value = Factory calibration value.
Normal display