16
Operation
K0394 Issue 3 - [EN] English
Menu: Set FSO
low/high registers
Use the FSO low/high registers to set a different range for the
functions that follow: analog display, % indication, low/high alarm.
Initially, these register values are set to the factory calibration
values. Example:
Calibrated range: 0.7 bar (10 psi) gauge.
Selected units: mbar
Table 8 gives the permitted FSO values.
Table 8: Permitted FSO values
FSO low/high registers -
Set-up and use
Menu: Set the menu option to the FSO low register (refer to menu
operations, page 5). Use these steps to set an applicable value in
the permitted range, refer to Table 8 above:
3.
Repeat steps
1
&
2
for each digit and for the decimal point.
If an incorrect value is entered, the value resets to the nearest
permitted value, refer to Table 8 above.
To accept or change the new value, repeat steps
1
to
3.
4.
If necessary, repeat the procedure for the FSO high register.
FSO low
FSO high
0 bar (0 psi)
700 mbar (10 psi)
Range
Permitted FSO values
All ranges: a, sg
0 to 105% FS
g: 0.7 bar (10 psi)
-0.7 bar (-10 psi) to 105% FS
g:
≥
2 bar (30 psi)
-1 bar (-15 psi) to 105% FS
All ranges
FSO low < FSO high
FSO low
1
2
FSO low
0 to 9, or -
0 to 9
Press
To cancel the new value.