[EN] English - K0394 Issue 3
Operation
15
Lock code - Set-up and
use
Menu: Set to
On (
refer to menu operation, page 5). Use the steps
that follow to set a new code.
3.
Complete the lock code, repeat steps
1
&
2
for each digit.
4.
To change the menu options, the display shows: L - - -
5.
Enter the applicable code.
To reset the code to the factory code, do a restore operation. Refer
to menu operations, page 5.
Menu: Set scan rate
This function sets the rate that the DPI 104 uses to take pressure
samples from the internal sensor.
The nominal update rate for the display is 2 Hz. The up-date rate for
the maximum/minimum function and the switch function is greater
than or equal to the specified scan rate.
Note: Increasing the scan rate, increases the power consumption.
Scan rate - Set-up and
use
Menu: Set an applicable value in the range 2 to 10 Hz, refer to menu
operation, page 5.
Menu: Monitor
external IDOS
Use this function to read the pressure from an external IDOS UPM.
All the other DPI 104 pressure functions are available but not the
calibration function.
Example: Set tare (or set zero), Monitor maximum/minimum.
This function does not supply power to the IDOS UPM and needs
accessories B and C, refer to Table 3, page 4.
Monitor external IDOS -
Set-up and use
1. Connect the DPI 104 (Figure B2 (d)).
2. Menu: Set to
On
(refer to menu operation, page 5).
Lock
1
2
Lock
Digit = 0 or 9
Digit = 0 to 9