External software can be used with the DPI
104 (Table
3
accessories
A or D, page 4). If SiCal
PRO (accessory D) is used, all the
menu commands and display data are available.
Note: To use SiCal
PRO, it must have the serial number of each unit
to be used with (DPI
104, IDOS
UPM, PC6-IDOS).
Set-up a DPI 104
network
Set-up a network of up to 99 units in series (‘daisy chain’). Figure 4
shows the electrical connections to do this, refer to Table 6 page 8.
Figure 4:
Connections for a DPI 104 network
Error codes
Table 9: Error codes/indications
This symbol appears when the DPI 104 transmits or
receives data.
DPI 104 (1)
Rx
GND
(+)
PC: 9-way, D-type
connector
12-24 Vdc
DPI 104 (2)
GND
Tx
Code
Description
Action
E0001
Incorrect unlock code.
Use the correct code.
E0002
The tare facility is not available
because the menu lock is
On
and
the lock code < 500.
Change the menu configuration.
E0004
Start-up error.
Do a restore operation (refer to
menu operation, page 5).
E0005
External IDOS UPM not found.
Make sure that all the related
equipment and connections are
serviceable.
E0006
Incorrect calibration access code.
Use the correct code.
E0007
The power supply voltage is too low
to do a calibration.
Use an external power supply or
replace the battery.
E0009
Unable to supply the specified Vout. Example:
Low battery. Use an external power
supply or replace the battery.
Bad connection. Make sure that all
the related equipment and
connections are serviceable.
OLoAd
Applied pressure
≥
110% FS.
Reduce the pressure.
99999/
-9999
There are not enough digits in the
main display to give the correct
pressure value.
Change the measurement units.