background image

DigitalFlow™ XGS868i Programming Manual

iii

Preface

Information Paragraphs

Note:

These paragraphs provide information that provides a deeper understanding of the situation, but is not 
essential to the proper completion of the instructions.

IMPORTANT:

These paragraphs provide information emphasizing instructions which are essential to proper setup of 
the equipment. Failure to follow these instructions carefully may cause unreliable performance.

Safety Issues

Auxiliary Equipment

Local Safety Standards

The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards, 
regulations, or laws applicable to safety.

Working Area

WARNING!

Indicates a potentially hazardous situation which can result in serious personal 

injury or death, if it is not avoided.

CAUTION!

Indicates a potentially hazardous situation which can result in minor or moderate 

injury to personnel or damage to the equipment, if it is not avoided.

HIGH VOLTAGE!

This symbol indicates the presence of high voltage. It calls your attention to 

situations or operations that could be dangerous to you and other persons operating the 
equipment. Read these messages and follow the instructions carefully.

WARNING!

It is the responsibility of the user to make sure all local, county, state and national 

codes, regulations, rules and laws related to safety and safe operating conditions are met for 
each installation.

WARNING!

Auxiliary equipment may have both manual and automatic modes of operation. As 

equipment can move suddenly and without warning, do not enter the work cell of this equipment 
during automatic operation, and do not enter the work envelope of this equipment during 
manual operation. If you do, serious injury can result.

WARNING!

Make sure that power to the auxiliary equipment is turned OFF and locked out 

before you perform maintenance procedures on the equipment.

Summary of Contents for DigitalFlow XGS868i

Page 1: ...GE Measurement Control Gas Analysis 910 196P Rev G February 2015 DigitalFlow XGS868i Panametrics Steam Ultrasonic Mass Flow Transmitter 1 2 Channel Programming Manual ...

Page 2: ......

Page 3: ...neral Electric Company All rights reserved Technical content subject to change without notice DigitalFlow XGS868i Panametrics Steam Ultrasonic Mass Flow Transmitter 1 2 Channel Programming Manual 910 196P Rev G February 2015 ...

Page 4: ... no content intended for this page ...

Page 5: ...ituation which can result in minor or moderate injury to personnel or damage to the equipment if it is not avoided HIGH VOLTAGE This symbol indicates the presence of high voltage It calls your attention to situations or operations that could be dangerous to you and other persons operating the equipment Read these messages and follow the instructions carefully WARNING It is the responsibility of th...

Page 6: ...quipment WEEE take back initiative directive 2012 19 EU The equipment that you bought has required the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the ...

Page 7: ...ting Up Inputs 11 1 7 1 Zero Cutoff Value 11 1 7 2 Assume Saturation 11 1 7 3 Steam Input Type 11 1 7 4 Temperature Input 12 1 7 5 Base Temperature 12 1 7 6 Pressure Input 12 1 7 7 Base Pressure 13 1 7 8 Low Pressure Switch 13 1 7 9 Quality Input 13 1 8 Entering Setup Data 14 1 8 1 Setting Transducer Signal Parameters 14 1 8 2 Initializing Setup Parameters Default Setup 16 1 8 3 Setting Response T...

Page 8: ...e Data Logging Option Card 55 3 3 Logging via PanaView 55 3 4 Creating Meter Logs 56 3 4 1 Creating a Standard Meter Log 57 3 4 2 Creating an Error Log 59 3 4 3 Checking Log Memory 60 3 4 4 Stop Logging 60 3 5 Creating PC Logs 61 3 6 Viewing Meter Log Files 64 3 7 Viewing PC Log Files 66 Chapter 4 Printing Data 4 1 Data Types for Printing 69 Chapter 5 Clearing Data 5 1 Introduction 71 5 2 Clearing...

Page 9: ...ters 102 C 5 5 Entering Setup Parameters 106 C 6 Entering Data in the Global Menu 110 C 6 1 Entering Global System Data 111 C 6 2 Setting Up Inputs and Outputs 113 C 6 3 Entering Communications Data 125 C 7 Exiting the Site Edit Menu 131 C 8 Saving Site Data 132 C 8 1 Saving Current Site Data to the Meter 133 C 8 2 Saving New Site Data to the XGS868i 134 C 8 3 Saving a Site to the PC 134 C 8 4 Cle...

Page 10: ...Contents viii DigitalFlow XGS868i Programming Manual ...

Page 11: ... channels and Scroll to the desired measurement method Channelx SYSTM enter the individual channel parameters Channelx PIPE enter the pipe geometry and other parameters Channelx I O set up the inputs and outputs Channelx SETUP set the signal limits response times and mass flow status Global SYSTM Scroll to system units English or metric Global I O set up error handling option cards and the display...

Page 12: ...a create and save graph and log files display text output and graphs of live measurement data create custom templates for displaying text graph and log data interface with multiple GE Measurement Control instruments Although the actual displays differ somewhat the general procedures are the same for all three programming methods This chapter provides detailed programming instructions for use with ...

Page 13: ...p may be followed in sequence or the four arrow keys may be used to scroll through the prompt screens Figure 1 shows the front of the XMT868i with magnetic keypad and magnetic wand Figure 1 XGS868i Magnetic Keypad and Wand IMPORTANT The XGS868i s keypad permits programming of the instrument through the glass faceplate without removing the cover Thus all programming procedures may be performed whil...

Page 14: ...e prior key As a guide in following the programming instructions in this chapter the relevant portions of the Model XGS868i menu map have been reproduced in Figure 28 on page 77 and Figure 29 on page 78 Proceed to the following sections to enter data in the Channel or GLOBL menus IMPORTANT If the keypad has not been pressed for 10 minutes the XGS868i exits the Keypad Program and returns to display...

Page 15: ...Skan Only is the preferred technique for locating the acoustic signal and for high velocity measurements It is more robust in a noisy environment than the Measure technique Skan Measure is the preferred technique to use for low velocity measurements If Skan Only is selected at the above prompt the meter uses this technique exclusively However if Skan Measure is selected the meter uses Skan Only to...

Page 16: ...nits for the flow rate display and press Enter 3 Scroll to the desired number of Vol Decimal Digits digits to the right of the decimal point in the volumetric flow rate display and press Enter 1 5 3 Selecting Totalizer Units 1 Scroll to the desired Totalizer Units for the totalized flow rate display and press Enter Available units are listed in Table 1 2 Scroll to the desired number of Tot Decimal...

Page 17: ...ss Enter The available units for this prompt are determined by the selection made at the System Units prompt 5 Scroll to the desired number of Mass Decimal Digits digits to the right of the decimal point in the totalized mass flow rate display and press Enter Procedure Options After completing the steps on the previous page the XGS868i returns to the Channel PROGRAM window Do one of the following ...

Page 18: ... Special transducers have no engraved number on the head and are rarely used Examine the transducer head carefully for a number If you entered the number for a standard transducer proceed to the Pipe OD prompt in step 5 If you entered the number for a special transducer proceed to step 3 below 1 6 1 Special Transducers Note For special transducers GE Measurement Control will supply a transducer da...

Page 19: ...arrow keys to enter the known Pipe Wall Thickness in inches or mm and press Enter If the pipe wall thickness is not available look up the value in a table of standard pipe size data which can be found in the Sound Speeds and Pipe Size Data manual 914 004 IMPORTANT Because the units cannot be independently chosen for this parameter the value must be entered in the same units used for the pipe OD Ci...

Page 20: ...luid Soundspeed in feet per second in the gas to be measured and press Enter 11 Scroll to the appropriate choice to indicate whether you want Reynolds Correction and press Enter If Off is selected go to Step 12 If On is selected the program asks for the Kinematic Viscosity Use the arrow keys to enter the desired value and press Enter 12 Use the arrow keys to enter a value for the flow Calibration ...

Page 21: ... from 0 to 1 ft sec 0 to 0 30 m sec for the Zero Cutoff and press Enter The recommended setting is 0 0200 ft sec 0 0009 m sec 1 7 2 Assume Saturation This prompt permits you to use a single input for either temperature or pressure to measure mass flow The User Program contains a table that correlates pressure and temperature for saturated steam If the meter can assume the steam is always at or nea...

Page 22: ...dures would be used to set up Input B 1 7 5 Base Temperature 6 Use the arrow keys to enter the Base Temperature and press Enter The ratio of this value to the actual temperature is used to calculate the standard mass flow rate If you selected No at Assume Saturation proceed to Step 7 below If you selected Pressure as the Steam Input Type proceed to Step 7 below If you selected Temperature as the S...

Page 23: ...s a quality value of 1 0 indicates 100 gas 12 At the Quality Input prompt scroll to a fixed steam quality value or to set up the option card in Slot 1 that will supply the live steam quality input and press Enter Note If Slot 1 contains an activated option card with an analog input that has been programmed as Special Slot 1 appears as an option at the above prompt Normally a fixed value of 1 0 100...

Page 24: ...fecting the transducer signal For example the programmed signal strength low limit may be used to determine the trigger point for an alarm CAUTION The Signal default settings are suitable for most applications Consult the factory before changing any of these parameters Only after consulting the factory complete the following steps to program the signal parameters 1 In the Keypad Program scroll to ...

Page 25: ...y exceeds the programmed VELOCITY HIGH LIMIT value See Chapter 2 in the Service Manual for a discussion of error codes Acceleration Limit 0 to 250ft s 0 to 76m s 50 ft s 15 m s The E6 CYCLE SKIP error message appears when the calculated fluid velocity changes by more than the programmed ACCELERATION LIMIT value from one reading to the next See Chapter 2 in the Service Manual for a discussion of er...

Page 26: ...ch value DO NOT change this value unless advised by the factory Shifts 0 to 10 3 The number of shifts corresponds to the actual number of transmits per cycle number of signals added together in one direction to produce an averaged signal for one interrogation of the fluid and need only be changed if the environment is very noisy or the acoustic signal is weak A Divisor 0 1 to 10 2 5 The Divisor us...

Page 27: ...y GE Measurement Control personnel 1 In the Keypad Program scroll to PROG and press Enter 2 In the PROG menu scroll to CH1 or CH2 and press Enter 3 In the Channel PROGRAM menu scroll to Setup and press Enter 4 Scroll to AVRG and press Enter 5 Scroll to the Response Time in number of readings and press Enter For best results select a value of 30 to ensure the most stable signal Procedure Options Af...

Page 28: ... press Enter 3 In the Channel PROGRAM menu scroll to Setup and press Enter 4 Scroll to ADVAN Press Enter 5 Scroll to MULTK and press Enter 6 Scroll to ON to activate or to OFF to deactivate multiple K factors and press Enter Note If you selected OFF go to Procedure Options on the next page 7 Scroll to the Custom Type of K factor desired CstV velocity or CstR Reynolds Press Enter If you selected Cs...

Page 29: ... calculate mass flow 1 In the Keypad Program scroll to PROG and press Enter 2 In the PROG menu scroll to CH1 or CH2 and press Enter 3 In the Channel PROGRAM menu scroll to Setup and press Enter 4 Scroll to ADVAN Press Enter 5 Scroll to MASS and press Enter 6 Scroll to Yes to activate or to No to deactivate Static Density and press Enter Note If you scroll to No go to Procedure Options below 7 Ente...

Page 30: ...um difference or average readouts data from the Global System submenu is used Any conflicting data entered in the Channelx System submenu is overridden The following submenus are included in the Global menu SYSTM use to specify the units of measure used in calculations see the next page I O used to set up error handling and to configure analog inputs and outputs see page 23 COMM used to set up the...

Page 31: ... the desired Atmospheric Pressure press Enter and proceed to the next step 7 Do one of the following 1 Channel meter go to Procedure Options on page 22 2 Channel meter proceed to Selecting Volumetric Units below 1 9 1a Selecting Volumetric Units 1 Scroll to the desired Volumetric Units for the flow rate display and press Enter Available units are listed in Table 5 2 Scroll to the desired Time unit...

Page 32: ...t are determined by the selection made at System Units see Table 6 2 Scroll to the desired Mass Flow Time units for the mass flow rate display and press Enter 3 Scroll to the desired number of Mdot Decimal Digits digits to the right of the decimal point in the mass flow rate display and press Enter 4 Scroll to the desired Mass Totalizer units for the totalized mass flow rate display and press Ente...

Page 33: ...ion made at the above prompt Remember to record all programmed data in Appendix B Data Records 1 9 2a Selecting Error Handling This menu option permits programming of the manner in which the Model XGS868i s outputs respond during an error condition See Chapter 2 Error Codes in the Service Manual for a discussion of the built in error codes To access this submenu Note The 2 Channel meters have an a...

Page 34: ...ror handling occurs when either channel goes into error Procedure Options After completing the above steps the XGS868i returns to the Global I O window Do one of the following To continue programming refer to Appendix A Menu Maps to navigate to the desired menu To leave the Keypad Program press Escape four times Table 7 Error Response Options Option Output Response Totalizer Response HOLD Holds th...

Page 35: ... page 29 Alarm Relays see page 30 Totalizer Outputs see page 32 Frequency Outputs see page 34 1 9 2c Analog Outputs While following the programming instructions refer to Figure 31 on page 80 Accessing the Output 1 In the Keypad Program scroll to PROG and press Enter 2 In the PROG menu scroll to GLOBL and press Enter 3 In the Global PROGRAM menu scroll to I O and press Enter 4 Scroll to OPTN and pr...

Page 36: ... A TOTL Displays the reverse totalized volume flow N A N A TIME Displays the total flow measurement time N A N A MDOT Displays the mass flow N A N A MASS Displays the forward totalized mass flow N A N A MASS Displays the reverse totalized mass flow N A N A SS up Displays the signal strength for the upstream transducer 50 75 50 or 75 SS do Displays the signal strength for the downstream transducer ...

Page 37: ... N A N A AcVOL Displays actual volumetric flow N A N A StVOL Displays standard volumetric flow N A N A z Displays the compressibility factor 0 1 0 or 1 Tu S2 Displays Skan transit time upstream N A N A Td S2 Displays Skan transit time downstream N A N A DT S2 Displays Skan Delta T N A N A Tu M2 Displays Measure transit time upstream N A N A Td M2 Displays Measure transit time downstream N A N A DT...

Page 38: ... and press Enter 2 Scroll to the desired input Measurement and press Enter Do one of the following If Off was selected go to Procedure Options below If Pressure or Temperature was selected proceed to Step 5 If Special was selected to set up the input as a live special input proceed to the next step 3 Use the arrow keys to enter a Name for the input and press Enter 4 Use the arrow keys to enter a U...

Page 39: ...te Input letters A B etc correspond to input wiring terminal numbers from top to bottom e g Input A corresponds to pins 1 2 and 3 Input B corresponds to pins 4 5 and 6 etc Setting Up the RTD Input 1 Enter a Label for the RTD input and press Enter 2 Scroll to the desired input type and press Enter Do one of the following If Off was selected go to Procedure Options below If Temperature was selected ...

Page 40: ...nal numbers from top to bottom e g Output A corresponds to pins 1 2 and 3 Output B corresponds to pins 4 5 and 6 etc Selecting the Alarm Type 1 Scroll to the desired Alarm type and press Enter 2 Do one of the following If you selected Off the XGS868i returns to the I O window Go to Procedure Options on the next page If you selected High Low or Fault proceed to the next step 3 Scroll to the appropr...

Page 41: ...ement units that appear in these prompts are those selected in the Global System window earlier in this section 8 Enter a value for the Trigger Point of the alarm and press Enter Go to Procedure Options below 9 Scroll to the type of Error flow non flow or both that will trigger the fault alarm and press Enter Procedure Options After completing the previous steps the XGS868i returns to the I O wind...

Page 42: ...2 In the PROG menu scroll to GLOBL and press Enter 3 In the Global PROGRAM menu scroll to I O and press Enter 4 Scroll to OPTN and press Enter 5 Scroll to Slot 1 and press Enter 6 Scroll to the desired Output and press Enter 7 Do one of the following If you selected Off the XGS868i returns to the I O window Go to Procedure Options on the following page If you selected TTLZR and are using a 1 Chann...

Page 43: ... sec for the frequency of the totalizer pulses and press Enter Note A complete pulse consists of equal amounts of ON and OFF times Choose a value that is compatible with the frequency counter to be used 4 Use the arrow keys to enter a value for the number of measurement units represented by each pulse and press Enter Procedure Options After completing the above steps the XGS868i returns to the I O...

Page 44: ...L and press Enter 3 In the Global PROGRAM menu scroll to I O and press Enter 4 Scroll to OPTN and press Enter 5 Scroll to Slot 1 and press Enter 6 Scroll to the desired Output and press Enter 7 Do one of the following If you selected Off the XGS868i returns to the Input Output window Go to Procedure Options on the following page If you selected Frequency and are using a 1 Channel meter proceed to ...

Page 45: ...this section 3 Use the arrow keys to enter a value for the Low end of the frequency output range and press Enter Note A complete pulse consists of equal amounts of ON and OFF times Choose a value that is compatible with the frequency counter to be used 4 Use the arrow keys to enter a value for the High end of the frequency output range and press Enter 5 Use the arrow keys to enter a value between ...

Page 46: ...ds using PanaView software with this link Use the COMM submenu to set the communications port and MODBUS parameters While programming the meter refer to Figure 31 on page 80 1 9 3a Setting Up the Serial Port Use the steps below to configure the communications port 1 In the Keypad Program scroll to PROG and press Enter 2 In the PROG menu scroll to GLOBL and press Enter 3 In the Global PROGRAM menu ...

Page 47: ...B LSB 1 Register 16 bit integer Integer 32 bit IntegerI MSB LSB LSB LSB 2 Register 32 bit integer Floating Point FP EXP MAN MAN MAN 2 Registers 32 bit IEEE floating point number Use the steps below to configure MODBUS communications IMPORTANT The XGS868i MODBUS communications settings chosen in the next four steps must match those of the MODBUS control system 1 Scroll to the desired MODBUS Baud Ra...

Page 48: ...imal places Size in Bytes 1 0 1 Clear Ch1 Totalizers 2 16 bit signed 2 2 Not Used 2 16 bit signed 3 4 Velocity 2 4 2 16 bit int 5 8 Act Volumetric Q DIGITS 4 IEEE 32 bit 7 C Std Volumetric Q DIGITS 4 IEEE 32 bit 9 10 Fwd Totals T DIGITS 4 2 16 bit int 11 14 Rev Totals T DIGITS 4 2 16 bit int 13 18 Tot Digits 0 2 14 1A Mass Flow M DIGITS 4 IEEE 32 bit 16 1E Fwd Mass Totals MT DIGITS 4 2 16 bit int ...

Page 49: ...Mass Tot Digits 0 2 21 28 Ch1 Timer 2 4 2 16 bit int 23 2C Ch1 Error Code 0 2 24 2E Ch1 Sound Speed 3 4 2 16 bit int 26 32 Ch1 Density 4 4 2 16 bit int 28 36 Ch1 Sig Strength Upstream 1 4 2 16 bit int 30 3A Ch1 Sig Strength Downstream 1 4 2 16 bit int 32 3E Ch1 Temperature 2 4 2 16 bit int 34 42 Ch1 Pressure 3 4 2 16 bit int 36 46 Ch2 Velocity 2 4 2 16 bit int 38 4A Ch2 Act Volumetric Q DIGITS 4 I...

Page 50: ...t int 65 80 Ch2 Temperature 2 4 2 16 bit int 67 84 Ch2 Pressure 3 4 2 16 bit int 69 88 Avg Velocity 2 4 2 16 bit int 71 8C Avg Act Volumetric Q DIGITS 4 IEEE 32 bit 73 90 Avg Std Volumetric Q DIGITS 4 IEEE 32 bit 75 94 Avg Fwd Totals T DIGITS 4 2 16 bit int 77 98 Avg Rev Totals T DIGITS 4 2 16 bit int 79 9C Avg Tot Digits 0 2 80 9E Avg Mass Flow M DIGITS 4 IEEE 32 bit 82 A2 Avg Fwd Mass Totals MT ...

Page 51: ... GLOBL and press Enter 3 Scroll to SECUR and press Enter 4 At the Lock Out prompt scroll to UNlck to unlock the system and return to the initial PROGRAM menu or scroll to LOCK to lock the system Note If the system was unlocked at the above prompt the following steps do not appear 5 Enter the current or default Password and press Enter 6 At the Edit Password prompt scroll to NO and press Enter to l...

Page 52: ...Chapter 1 Programming Site Data 42 DigitalFlow XGS868i Programming Manual ...

Page 53: ... contrast of the LCD may be adjusted for optimum viewing Proceed to the appropriate section for instructions and refer to Figure 30 on page 79 of Appendix A Menu Maps 2 2 1 Adjusting LCD Contrast The LCD contrast may be adjusted through the Keypad Program to suit individual needs 1 Press Escape Enter Escape 2 In the Keypad Program window scroll to CNTRS and press Enter 3 Scroll to DARKN or LITEN a...

Page 54: ...Scroll to I O and press Enter 6 Scroll to LCD and press Enter 7 The window now asks for the of LCD Parameters Scroll to the desired number from OFF through 1 4 and KEY and press Enter The OFF setting switches the measurement display off while the KEY setting enables users to change the measurement display via the arrow keys without accessing the Keypad Program If you select KEY To view a different...

Page 55: ...faults to the nearest selectable item in the parameter list The previous two prompts repeat until all of the specified of LCD Parameters have been set up When all of the display parameters have been set up the meter returns to the Global I O window To leave the Keypad Program press the Escape key three times After leaving the Keypad Program the XGS868i will reset itself and will begin to display t...

Page 56: ...S Serial Communications 916 054 You must also install PanaView as discussed in the PanaView User s Manual 910 211 and in Appendix C Programming the XGS868i via PanaView 1 Power up the XGS868i and wait until it has initialized 2 Launch PanaView and wait until it initializes the XGS868i and displays the basic PanaView window 3 As described in Appendix C open the Meter Browser window and highlight th...

Page 57: ...e Output menu and click on the Text Display option see Figure 2 on page 46 Note The Text Display window that appears after Step 2 is actually stacked on top of any previously opened windows such as the Meter Browser window 3 Using the Window menu arrange the open windows in the desired format For this discussion Figure 3 shows the Text Display window in its maximized full screen size Figure 3 The ...

Page 58: ...sed by clicking on the lower X control button at the far right of the menu bar IMPORTANT If you click on the upper X control button at the far right of the PanaView title bar you will exit PanaView completely 2 3 1a Displaying Multiple Process Parameters The procedure for displaying a single process parameter in a text screen may be repeated to simultaneously display multiple process parameters To...

Page 59: ...trument and display it graphically in a new format complete the following steps 2 3 2a Setting Up a Graph 1 Perform Steps 1 3 on page 46 2 Pull down the Output menu and click on the Graphing New option see Figure 2 on page 46 Note The Graph window is stacked on top of any previously opened windows such as the Meter Browser window 3 Using the Window menu arrange the open windows in the desired form...

Page 60: ...hed for any other active meter channels 3 To begin plotting data for the selected parameter s enter a time Interval for data sampling in the text box provided the default value is 5 seconds 4 Click on the Plot Continuous control button to begin graphing the selected process parameter s as a function of time at the interval specified in Step 3 5 While PanaView is graphing the data the following act...

Page 61: ...GS868i users to read and plot transducer signals from the XGS868i 1 From the New Meter Browser highlight the XGS868i 2 Right click on the highlighted XGS868i and click on the Properties option as shown in Figure 5 Figure 5 The Properties Option in the New Meter Browser The Properties window opens as shown in Figure 6 Figure 6 The Properties Window ...

Page 62: ... down menu and click on the desired channel After a moment the Properties window appears similar to Figure 7 Figure 7 Active Signal Options 4 To select a different signal type open the signal menu at the right shown here with Raw Upstream highlighted and click on the desired signal 2 3 3b Plotting Transducer Signals To plot the selected signal click on Plot A graphical window opens as shown in Fig...

Page 63: ...ual 53 Chapter 2 Displaying Data 2 3 3c Saving Transducer Signals To save the raw signal click on Save A window opens similar to Figure 9 Enter the desired name and click on Save to save the signal as a text file Figure 9 Save As Window ...

Page 64: ...Chapter 2 Displaying Data 54 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 65: ...rage on the PC s hard drive This chapter explains how to use the Model XGS868i s data logging capability with PanaView 3 2 The Data Logging Option Card In order to log data the Model XGS868i must be fitted with an optional data logging option card in Slot 2 See Chapter 4 Parts Replacement of the Service Manual for instructions on installing the card 3 3 Logging via PanaView PanaView is capable of ...

Page 66: ...in PanaView expand the network tree and open the Edit Functions option described in Appendix C Programming the XGS868i via PanaView The menu will appear similar to Figure 10 Figure 10 The Edit Functions Menu for an XGS868i with a Logging Card 2 Double click on the Log Edit Menu as shown in Figure 10 The window appears similar to Figure 11 Figure 11 The Log Edit Menu Window ...

Page 67: ...er of Variables you wish to log Type in the desired number from 1 to 6 and click Next Item f Double click on the desired Channel for the first variable g Scroll to and double click on the desired Measurement to be logged h Repeat steps f and g for the specified number of variables If you select a forward or reverse total PanaView asks if you want to Reset Totals to 0 Double click on Yes or No i Pa...

Page 68: ...number in the right window and click Next Item or simply click Next Item to step through the parameters If you select Timed PanaView asks for the Log Time duration Scroll to and double click on the desired length of time from 10 minutes to 24 hours Then go to step m l PanaView indicates a given End Date Double click on OK to agree to the given day on Today to start the log that day or on Edit to c...

Page 69: ... total PanaView asks if you want to Reset Totals to 0 Double click on Yes or No h PanaView asks Is Log Circular Double click on Yes or No i PanaView indicates a given StartTime Double click on OK to agree to the given time on Now to start the log immediately or on Edit to change the starting time If you select Edit PanaView asks first for the desired Hour then for the Minute and finally for the Se...

Page 70: ...ge the starting date If you select Edit PanaView asks first for the desired Year then for the Month and finally for the Day For each parameter type the desired number in the right window or scroll to and click on the desired month and click Next Item or simply click Next Item to step through the parameters m Scroll to and double click on the desired Time Increment from 5 seconds to 24 hours PanaVi...

Page 71: ...s in the Logging Option You have two choices Click on the New PC Log option and proceed to step 5 or Click on the Log Browser option The dialog box shown in Figure 14 appears Figure 14 Computer Selection Note If no log files have yet been created or if a log file has not yet been selected from the list all of the option buttons except for Cancel in the Log Browser dialog box are unavailable 2 Doub...

Page 72: ...click on PC Logs to advance to the dialog box shown in Figure 16 Figure 16 Create a PC Log 4 At the dialog box shown in Figure 16 click on the Create option button to create a new meter log 5 Enter a log file name Sample in this case in the dialog box shown in Figure 17 Click on the OK option button Figure 17 Log Name Dialog Box ...

Page 73: ...he network tree as far as necessary and double click on the process parameter s to be logged As they are selected the parameters are listed in the lower right pane of the PC Log window make sure that the Log Items tab in this pane is selected Note To add all the channel parameters to a given log right click on the desired channel A pop up button Add Group to Log appears Click on the button and Pan...

Page 74: ...appear similar to Figure 19 Figure 19 An Expanded Meter Logs Option 2 Double click on the desired log A window opens for the individual log as shown in Figure 20 Figure 20 Data Window for an Individual Meter Log Click on the Log Items tab see Figure 20 for a list of the parameters specified for logging Click on the Contents tab see Figure 21 on page 65 for a list of all the data points logged to d...

Page 75: ...tput section in Chapter 2 of the PanaView User s Manual Remember that only two parameters per channel may be graphed and that the same two parameters must be graphed for each channel of a multi channel graph display When you have finished viewing the graph click on the Close option button to close the Graph window and leave the log running To review log details click on the Contents tab The window...

Page 76: ...g name to open the log or Pull down the Output menu see Figure 13 on page 61 and click on the Log Browser option Advance to the PC Logs dialog box as described in previous sections A screen similar to that shown in Figure 23 appears Figure 22 The Expanded PC Logs Option Figure 23 Selecting a Log File from the Log Browser Notice that each of the log files in Figure 22 and Figure 23 has a traffic li...

Page 77: ...d to date Note The graph is displayed in its own window which is opened on top of the PC Log window For instructions on using the Graph Log window refer to the Graphing Output section in Chapter 2 of the PanaView User s Manual Click on the Refresh option button to update the information shown on the Contents tab and in the Graph Log window Any data logged since the last use of the Refresh option b...

Page 78: ...Chapter 3 Logging Data 68 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 79: ...nal PanaView software Note See Chapter 1 Installation of the Startup Guide for instructions on wiring the RS232 serial port For additional information refer to the EIA RS Serial Communications Manual 916 054 After making the hardware connections and installing PanaView software the following data may be output to a printer connected to the personal computer live data in numeric or graphical format...

Page 80: ...Chapter 4 Printing Data 70 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 81: ...cted option are thoroughly understood before proceeding 5 2 Clearing the XGS868i s Memory If the XGS868i s available memory becomes nearly full it may be necessary to purge some or all of the existing data from memory before any additional data can be stored In order to accomplish this task proceed to the appropriate section for step by step instructions Note Depending on the current configuration...

Page 82: ...e Log Browser option as shown in Figure 24 Figure 24 The Log Browser in the Logging Option 2 Double click on the computer name to move to the network level 3 Double click on PC Logs to access the stored PC logs or on the communication port name and the meter name to access the stored meter logs In either case the Log Browser appears similar to Figure 25 Figure 25 Log Browser Window 4 Highlight the...

Page 83: ...i asks Reset Totals Scroll to NO or YES 2 and press Enter at the desired choice The XGS868i totalizers are reset to 0 and the meter returns to the Keypad Program Press Escape to return to the data display To reset the totals from PanaView 1 From the New Meter Browser in PanaView highlight the XGS868i 2 Open the Edit menu and select the Properties option as shown in Figure 26 Figure 26 The Properti...

Page 84: ...Chapter 5 Clearing Data 74 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 85: ...DigitalFlow XGS868i Programming Manual 75 Appendix A Menu Maps Appendix A Menu Maps ...

Page 86: ...76 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 87: ... English PATH LENGTH P PIPE WALL inch feet mm m Metric English AXIAL LENGTH L Other Steam FLUID TYPE FLUID SOUNDSPEED On Off REYNOLDS CORRECTION CALIBRATION FACTOR KINEMATIC VISCOSITY VOLUMETRIC UNITS CHANNEL LABEL SITE CHANNEL MESSAGE Metric English ACF KACF MMACF SCF KSCF MMSCF ACM KACM MMACM SCM KSCM MMSCM SEC MIN HR DAY VOLUMETRIC TIME 0 1 2 3 VOL DECIMAL DIGITS TOTALIZER UNITS 0 1 2 3 TOTAL D...

Page 88: ...SWITCH OF PEAK SKAN T OFFSET DELTA T OFFSET AMP DISCRIM HIGH AMP DISCRIM LOW ACCELERATION LIMIT VELOCITY HIGH LIMIT VELOCITY LOW LIMIT SOUNDSPEED LIMIT CORR PEAK LIMIT SIGNAL LOW LIMIT Default Setup V averaging Advanced Features NO YES 1 2 5 30 60 STATS 10 Signal Defaults RESPONSE TIME Multiple K Factors Mass Flow Calculation Activate Multi K Fact Mass from Static Dens No Yes CstV CstR NO YES Kin ...

Page 89: ...M KSCM MMSCM BAUD RATE 600 1200 See Figures 28 and 29 Eng METER MESSAGE Metric PRESSURE UNITS SYSTEM UNITS PSIa PSIg BARa BARg Kpaa Kpag Metric English ATMOSPHERIC PRESSURE 1 CHANNEL UNITS 2 CHANNEL UNITS VOLUMETRIC UNITS LCD HOLD LOW HIGH HHIGH OTHER YES NO 2 PATH ERROR HANDLING OF LCD PARAMS SUM DIF AVE CH1 CH2 TOTL TOTL TIME MDOT VEL VOLUM SS up SS do SNDSP T up MASS MASS REYN Tot K PEAK Qup Td...

Page 90: ... SUM DIF AVE TOTL TOTL MASS MASS UNITS PULSE MINIMUM ON Measurement Name Totalizer Outputs Slot 1 Outputs C D A B TEMP OFF Input measurement LABEL RTD Inputs Low High C D A B FREQ OFF CH1 CH2 SUM DIF AVE Measurement Name Slot 1 Outputs TOTL TOTL TIME MDOT VEL VOLUM SS up SS do SNDSP T up MASS MASS REYN Tot K PEAK Qup Tdown DELTA AMPdn CNTup CNTdn P up Qdown AMPup PRESR Z AcVOL StVOL P dn TEMP Tsup...

Page 91: ...11 TA 2 Totalizer Outputs 2 Standard Alarms 1223 02 OI 2 Current Inputs 1430 04 OR 2 RTD Inputs 1223 04 TI 2 Current Inputs 2 Totalizer Outputs 1473 05 TR 2 RTD Inputs 2 Totalizer Outputs 1223 06 FI 2 Current Inputs 2 Frequency Outputs 1473 07 FR 2 RTD Inputs 2 Frequency Outputs 1223 08 AI 2 Current Inputs 2 Standard Alarms 1473 10 AR 2 RTD Inputs 2 Standard Alarms 1225 13 CO 2 Current Outputs 122...

Page 92: ... Whenever an option card is installed or changed in the Model XGS868i flow transmitter record the type of card and any additional setup information in the appropriate row of Table 15 Table 15 Option Cards Installed Slot Type of Option Card Additional Setup Information 0 Analog Outputs A B 1 2 ...

Page 93: ...c Digits Mass Dec Digits Totalizer Units Global Input Output Error Handling Error Handling 2 Path Error No Yes Global Communications Port Meter Address MOD Parity Baud Rate MOD Stop Bits MOD Baud Rate MOD Address Channel Status Channel 1 Channel 2 Channel Status Off Burst Channel Status Off Burst Measure Mode Skan S M Measure Mode Skan S M Channel System Channel Label Channel Label Site Channel Ms...

Page 94: ...ation Factor Calibration Factor Channel Input Output Zero Cutoff Zero Cutoff Assume Sat No Yes Assume Sat No Yes Steam Input Type Temp Pressure Steam Input Type Temp Pressure Temp Input Temp Input Base Temp Base Temp Pressure Input Pressure Input Base Pressure Base Pressure Low Press Switch No Yes Low Press Switch No Yes Pressure Limit Pressure Limit Quality Input Quality Input Channel SETUP V Ave...

Page 95: ...ing Manual 85 Appendix B Data Records Channel SETUP Advanced Features Multi K Factors cont K Factor Velocity K Factor K Factor Velocity K Factor 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 20 20 Table 16 Setup Data cont ...

Page 96: ...86 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 97: ... Channel Setup set the signal limits response times and activating mass flow Global System enter the system units Global I O set up error handling option cards and display Global Comm set the serial port and MODBUS parameters As a programming aid a complete set of menu maps for the PROG menu is included in Appendix D PanaView Menu Maps The specific figure numbers will be referenced throughout this...

Page 98: ...aunch PanaView as discussed in Chapter 3 Initial Setup of the PanaView User s Manual 2 Open the New Meter Browser window in the File menu and expand the network tree Then highlight the My Computer Name branch by clicking on it 3 Pull down the Edit menu by clicking on it in the menu bar 4 Click on the New menu option to select it and a submenu opens with two choices on it see Figure 32 Figure 32 Th...

Page 99: ...M Port Type menu and click on the desired type or on TCP IP if the XGS868i is using an Ethernet connection Note If you select TCP IP the menu changes Proceed to the next page 8 Select any suitable available baud rate A baud rate of 19 200 is appropriate for almost all applications However if you experience periodic communication reliability problems you may wish to consider lowering the baud rate ...

Page 100: ...ss If the IP address is unknown run the Device Discovery software utility All units found will be identified by their MAC address and the assigned IP address In the Port Number box enter 2101 the default value 3 Click OK to complete data entry IMPORTANT If you are using Ethernet communications be sure the XGS868i is set up with the default communications parameters 9600 baud no parity no handshaki...

Page 101: ...need to assign a static IP address to the XGS868i complete the following steps IMPORTANT To assign a static IP address the XGS868i must be connected to an Ethernet network with a dynamic address assignment mechanism DHCP first 1 Run Ethernet Device Discovery software and identify the XGS868i s current IP address 2 Open your Internet browser Internet Explorer Netscape or other and type the XGS868i ...

Page 102: ... on it and then open the Edit menu on the menu bar if the communication port is not highlighted first the New Meter option is not active in the Edit menu 2 Click on the New option in the Edit menu see Figure 35 Figure 35 New Option in the Edit Menu 3 After clicking on the New option the Meter menu option appears Click on this option to select it 4 The New IDM Meter screen shown in Figure 36 opens ...

Page 103: ...ization is successful the Meter Browser shows a listing similar to Figure 37 Figure 37 The Updated Network Tree However if the settings do not match or there is some other difficulty a screen appears similar to Figure 38 Figure 38 Communication Error Screen The screen offers the options of trying again or of canceling the operation Click on the desired choice and then on OK to confirm the choice o...

Page 104: ...complete at least the sections of this chapter pertaining to those three submenus Except for those three submenus it is not necessary to program the Model XGS868i flowmeter in any particular order Therefore the sections of this chapter need not be completed in sequence Enter the user program as described below and proceed immediately to any section of interest To begin entering data into the XGS86...

Page 105: ... Channel 1 Channel 2 or Global highlight and double click on the desired menu in the left pane For example clicking on Channel 1 in Figure 40 opens the list of options shown in the center pane 5 To enter a particular option a Highlight and double click on the desired option in the center pane Figure 42 on page 96 shows the first entry Transducer Number in the Pipe Parameters option The title above...

Page 106: ...unchanged As you step through the menu the bottom panel lists the current settings you have modified or left unchanged as shown in Figure 43 If you modify or step through more than five items a scroll bar at the right of the panel lets you review the earlier settings Figure 43 Site Edit Menu with Current Settings 6 When you have completed entering parameters in a given option click Exit Page to cl...

Page 107: ... The window now appears similar to Figure 44 Figure 44 The Status Option in the Channel Menu 3 Double click on Burst to activate the channel path Note Burst is automatically selected for a 1 Channel meter 4 Double click on one of the measurement methods described below Skan Only is the preferred technique for locating the acoustic signal and for high velocity measurements It is more robust in a no...

Page 108: ...w display 6 Double click on the desired number of Decimal Digits digits to the right of the decimal point in the volumetric flow display 7 Double click on the desired Totalizer Units for the totalized flow rate display listed in Table 17 8 Double click on the desired number of Decimal Digits digits to the right of the decimal point in the totalized flow rate display The program now varies dependin...

Page 109: ...al point in the mass flow rate display 4 Double click on the desired Mass Totalizer units for the totalized mass flow rate display listed in Table 18 5 Double click on the desired number of Mass Decimal Digits digits to the right of the decimal point in the totalized mass flow rate display PanaView returns to the Channel PROGRAM window shown in Figure 41 on page 95 Proceed to the Pipe option on th...

Page 110: ...ial transducers which have no engraved number on the head are rarely used Examine the transducer head carefully for a number The menu now varies depending on your selection in step 2 If you entered the number for a standard transducer proceed to the Pipe OD prompt in step 4 If you entered the number for a special transducer proceed to step 3 below C 5 3a Special Transducers 3 For special transduce...

Page 111: ...and or are included in the documentation supplied with the meter For on site transducer installations refer to Appendix C Measuring P and L Dimensions in the Startup Guide for instructions 7 Click on the appropriate Axial Length unit type in the center pane Then enter the axial length of the ultrasonic signal in the right pane and click Next Item C 5 3d Fluid Type 8 Double click on the appropriate...

Page 112: ...Zero Cutoff Enter a value from 0 to 1 ft sec 0 to 0 30 m sec for the zero cutoff and click Next Item The recommended setting is 0 1 ft sec 0 03 m sec C 5 4b Assume Saturation This prompt permits you to use a single input for either temperature or pressure to measure mass flow The User Program contains a table that correlates pressure and temperature for saturated steam If the meter can assume the ...

Page 113: ...r temperature the meter assumes you are using a fixed temperature 2 Proceed to one of the following sections If you selected Fixed proceed to Step 3 If you selected Slot 1 proceed to Step 4 3 Enter the known Fixed Temp process temperature and click Next Item The meter will accept values from 328 to 1 832 F 200 to 1 000 C Proceed to Base Temperature below 4 Select Input A or Input B and click Next ...

Page 114: ...tep 4 3 Enter the known Fixed process Pressure and click Next Item The meter will only accept values from 0 to 5 000 psia Proceed to Base Pressure below 4 Double click on Input A or Input B The inputs were labeled during setup Note The set up of Input A is used as an example Identical procedures would be used to set up Input B C 5 4g Base Pressure 1 Enter the Base Pressure and click Next Item The ...

Page 115: ...ve steam quality input and click Next Item Note If Slot 1 contains an activated option card with an analog input that has been programmed as Special Slot 1 appears as an option at the above prompt Normally a fixed value of 1 0 100 steam should be used for the steam quality value If there is no active option card for a special input the meter assumes you are using a fixed quality value If you selec...

Page 116: ...e Set Up submenu highlight and double click on the Set Up option in the center pane of the Channel menu Remember to record all programmed data in Appendix B Data Records C 5 5a The Signal Option Use this option to set the limits for the incoming signal and other parameters affecting the transducer signal For example the programmed signal strength low limit may be used to determine the trigger poin...

Page 117: ...ee Chapter 2 in the Service Manual for a discussion of error codes Acceleration Limit 0 to 250 ft s 0 to 76 m s 15 ft s 5 m s The E6 CYCLE SKIP error message appears when the calculated fluid velocity changes by more than the programmed ACCELERATION LIMIT value from one reading to the next See Chapter 2 in the Service Man ual for a discussion of error codes Amplitude Discriminator Low 0 to 100 14 ...

Page 118: ...rapid response to changes in flow rate After responding to the above prompt the meter returns to the Set Up option window M S Switch 0 to 250 μsec 50 μsec If the burst mode is set to Skan Measure S M the meter switches from Skan to Measure Mode when Delta T is less than the M S_Switch value DO NOT change this value unless advised to do so by the factory Shifts 0 to 10 3 The number of shifts corres...

Page 119: ... was selected the meter returns to the Advanced Features window If YES was selected proceed to Step 3 3 Double click on the desired Custom Type velocity or reynolds If you selected CstR Reynolds enter the desired Kinematic Viscosity and click Next Item 4 Double click on Yes to Edit the K factor Table or on No to retain the current K factor table and return to the Advanced Features window Note If t...

Page 120: ...r information that is not specific to any of the individual channels Information programmed via this menu is used to enter several general system parameters e g English or metric units For meters with 2 channels this menu is also used to compute parameters such as the sum difference or average of the channel 1 and channel 2 signals When calculating the SUM DIF or AVE readouts data from the Global ...

Page 121: ...ow appears similar to Figure 45 Figure 45 The System Option in the Global Menu 3 Double click on the desired System Units selection either metric or English The XGS868i will display all parameters and measurements in the designated units 4 Double click on the desired Pressure Units absolute or gauge and click Next Item a If gauge was selected enter the desired Atmospheric Pressure and click Next I...

Page 122: ...sted in Table C 5 above 2 Double click on the desired number of Tot Decimal Digits digits to the right of the decimal point in the totalized flow rate display 3 Do one of the following If Mass Flow is ON proceed to Selecting the Mass Flow Units below If Mass Flow is OFF the meter returns to the Global PROGRAM window Click Exit Page to return to the Site Edit Menu Note To activate mass flow refer t...

Page 123: ...ed entering selections in this option The program returns to the Global PROGRAM menu C 6 2 Setting Up Inputs and Outputs Set up the XGS868i s inputs and outputs via the I O submenu While following the programming instructions refer to Figure 55 on page 142 in Appendix D PanaView Menu Maps Remember to record all programmed data in Appendix B Data Records The I O submenu consists of the following op...

Page 124: ...step 3 on page 115 Table 23 Error Options Responses for Single Channel Meter Option Output Response Totalizer Response Hold Last Value Holds the last good reading Holds the last good reading and continues to totalize based on that reading Force Low Forces the outputs to the low set point Stops totalizing Force High Forces the outputs to the high set point Stops totalizing Force High High Forces th...

Page 125: ...ling or on No to disable this function Specific responses of the display and the totalizer to the two path error handling option available at the above prompt are listed in Table 25 After responding to the above prompt the meter returns to the Input Output option menu Table 25 2 Path Error Response Options Option Display Response Totalizer Response NO Displays the average of CH1 and CH2 regardless...

Page 126: ...e prompt Proceed to the appropriate section for programming instructions specific to the type of input or output selected above Analog Outputs Complete the following steps to set up the analog outputs for any slot refer to Figure 55 on page 142 1 Double click on the desired Output A or B Note The set up of output A is used here as an example Identical procedures would be used to set up output B 2 ...

Page 127: ...erence between the upstream and downstream signals N A N A Reynolds Displays the Reynolds number N A N A K Re MultiK KFactor K factor based on the Reynolds number 0 3 3 0 N A PEAK Displays the percentage of peak set to 50 by default N A N A UP Signal Q Displays signal quality for upstream transducer 800 800 DN Signal Q Displays signal quality for downstream transducer 800 800 UP Amp Discrim Displa...

Page 128: ...Displays Measure transit time upstream N A N A DN Transit M2 Displays Measure transit time downstream N A N A DeltaT M2 Displays Measure Delta T N A N A Vinst Displays the instantaneous velocity N A N A T Saturat Displays saturation temperature for steam at current pressure N A N A T Superht Displays actual temperature minus the saturation temperature Tactual Tsat 0 0 Fluid Dens Displays fluid den...

Page 129: ...another menu TEMP or PRESR If TEMP or PRESR was selected to set up input A as a live temperature or pressure input complete steps 4 and 5 below 4 Enter a value for the Base low end of the analog input range and click Next Item 5 Enter a value for the Full Scale high end of the analog input range and click Next Item After responding to the above prompt the meter returns to step 1 Click Exit Page to...

Page 130: ...s used here as an example Identical procedures would be used to set up the remaining RTD inputs 2 Enter a Label of up to eight characters for input A and click Next Item 3 Double click on Off to disable Input A and return to step 1 or on Temp to enable input A as a live temperature input 4 Enter a temperature value for the Zero low end of the analog input range and click Next Item 5 Enter a temper...

Page 131: ... 4 Double click on the desired Channel option See Table 28 for a description of the channel options available Note If FAULT was selected as the alarm type the next two prompts do not appear You have completed entering data in this option 5 Double click on the desired Measurement Name Table 27 on page 117 provides a description of the available parameters Note The measurement units that appear in t...

Page 132: ... Channel option See Table 28 on page 121 for a description of the channel options available at the above prompt 4 Double click on the desired Measurement Name as listed in Table 29 Note The measurement units that appear in these prompts are those selected in the Global System menu earlier in this section 5 Enter a value between 50 sec and 500 000 sec for Pulse On Time the frequency of the totalize...

Page 133: ...red Channel option See Table 28 on page 121 for a description of the channel options available at the above prompt 4 Double click on the desired Measurement Name Table 27 on page 117 provides a description of the available parameters Note The measurement units that appear in these prompts are those selected in the Global System menu earlier in this section 5 Enter a flow rate value for the Zero lo...

Page 134: ...EY The OFF setting switches the measurement display off while the KEY setting enables users to change the measurement display via the arrow keys without accessing the Keypad Program If you select KEY To view a parameter other than the one currently displayed press the or keys to scroll through the various parameters To scroll through the channel options on a two channel XGS868i press the or keys u...

Page 135: ... a Meter Address number between 1 and 254 and click Next Item The default number is 1 A meter address is only necessary for communication with the GE Measurement Control PanaView software See the PanaView User s Manual for more information IMPORTANT If the meter address or baud rate is changed communication with PanaView must be re established with the new address number 3 Double click on the desi...

Page 136: ...10 4 32 bit integer 8 CH1 Totals TOT_U Register 10 4 32 bit integer 10 CH1 T Digits none 0 2 16 bit integer 11 CH1 Totalizer Time sec 2 4 32 bit integer 13 2 CH1 Error Value none 0 2 16 bit integer 14 CH 1 SSUP none 1 4 32 bit integer 16 CH 1 SSDN none 1 4 32 bit integer 18 CH 1 SNDSP ft s or m s 0 4 32 bit integer 20 CH 2 Velocity ft s or m s 2 4 32 bit integer 22 CH 2 Volumetric VOL_U 4 IEEE 32 ...

Page 137: ...2 CH 1 Energy Digits none 0 2 16 bit integer 63 CH 1 TempS F or C 2 4 32 bit integer 65 CH 1 TempR F or C 2 4 32 bit integer 67 CH 1 TS TR F or C 2 4 32 bit integer 69 CH 1 DELTH Btu lb or J gm 2 4 32 bit integer 71 CH 2 Power Power_u 4 IEEE 32 bit 73 CH 2 Energy Energy_u Register 77 4 32 bit integer 75 CH 2 Energy Energy_u Register 77 4 32 bit integer 77 CH 2 Energy Digits none 0 2 16 bit integer...

Page 138: ...nd channel 2 if both channels out of error channel 1 value if channel 2 is in error channel 2 value if channel 1 is in error zero if both channels in error 4 Average Error Status 0 both in error 1 chan 2 in error 2 chan 1 in error 3 both ok 5 MODBUS baud rate 5 2400 6 4800 7 9600 6 MODBUS parity 0 none 1 odd 2 even 7 MODBUS stop bits 1 1 stop bit 2 2 stop bits 8 General Registers are written if co...

Page 139: ...nd channel 2 if both channels out of error channel 1 value if channel 2 is in error channel 2 value if channel 1 is in error zero if both channels in error 4 Average Error Status 0 both in error 1 chan 2 in error 2 chan 1 in error 3 both ok 5 MODBUS baud rate 5 2400 6 4800 7 9600 6 MODBUS parity 0 none 1 odd 2 even 7 MODBUS stop bits 1 1 stop bit 2 2 stop bits 8 General Registers are written if co...

Page 140: ... a default password 2719 and three spaces For increased security the default password should be changed Note Once the system has been locked it can only be unlocked by entering the password because access to the SECUR submenu is restricted While completing these instructions refer to the menu map in Figure 54 on page 141 1 In the Site Data Menu double click on the Global entry If the system is unl...

Page 141: ...s manual a new password should be entered If the password is ever lost contact the factory for assistance 6 Enter the New Password and click Yes Any combination of letters and numbers up to a total of 7 characters may be used as a password 7 Verify the New Password by re entering it and click Yes Be sure to record the new password in a safe place C 7 Exiting the Site Edit Menu After leaving the Gl...

Page 142: ...hrough PanaView users can store site file data in a PC and reload it into the XGS868i To save or reload site data via PanaView 1 In the New Meter Browser highlight the XGS868i 2 Right click on the highlighted XGS868i and select the Properties option as shown in Figure 49 Figure 49 The Properties Option The window now appears similar to Figure 50 Figure 50 The Properties Window for an XGS868i ...

Page 143: ... Figure 51 The Site File Operations Window C 8 1 Saving Current Site Data to the Meter To save existing site data to the meter 1 Select the radio button for Selected and highlight an existing site in the left pane 2 Then click on the Save Site to Meter button A screen opens similar to Figure 52 Figure 52 Site File Selection 3 Highlight the desired site and click Open PanaView sends the site to the...

Page 144: ...ns Highlight a site file with the desired settings and click Open The site file becomes the Working site in the meter with the desired settings C 8 3 Saving a Site to the PC To save a site to the PC 1 Highlight the desired site in the left pane See Figure 51 on page 133 2 Click Save Site to PC A window opens similar to Figure 52 on page 133 3 Enter the desired site name and click Save PanaView sav...

Page 145: ...the XGS868i on page 134 C 8 5 Saving Site Data in Text Form To store the data from a site file as a text file for display or printout 1 Highlight the site in the left pane See Figure 51 on page 133 2 Click on Save Site Print to PC 3 The Site File Selection window shown in Figure 52 on page 133 opens Enter the desired site name now with a prt suffix and click on Save PanaView displays a text versio...

Page 146: ...Appendix C Programming the XGS868i Using PanaView 136 DigitalFlow XGS868i Programming Manual ...

Page 147: ...DigitalFlow XGS868i Programming Manual 137 Appendix D PanaView Menu Maps for the XGS868i Appendix D PanaView Menu Maps for the XGS868i ...

Page 148: ...138 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 149: ...et mm m Metric English AXIAL LENGTH L Other Steam FLUID TYPE FLUID SOUNDSPEED On Off REYNOLDS CORRECTION CALIBRATION FACTOR KINEMATIC VISCOSITY VOLUMETRIC UNITS CHANNEL LABEL SITE CHANNEL MESSAGE Metric English ACF KACF MMACF SCF KSCF MMSCF ACM KACM MMACM SCM KSCM MMSCM SEC MIN HR DAY VOLUMETRIC TIME 0 1 2 3 VOL DECIMAL DIGITS TOTALIZER UNITS 0 1 2 3 TOTAL DECIMAL DIGITS Metric English MASS FLOW U...

Page 150: ...g Advanced Features NO YES 1 2 5 30 60 STATS 10 Signal Defaults RESPONSE TIME Multiple K Factors Mass Flow Calculation Activate Multi K Fact Mass from Static Dens No Yes CstV CstR NO YES Kin Viscosity Edit table Enter of K Factors Velocity Reynolds K Factor Repeat for the of K Factors entered previously Custom Type Fluid Density Edit Functions SITE EDIT MENU CALIBRATION TEST PAUSE MEASUREMENT MEAS...

Page 151: ...PRESSURE 1 CHANNEL UNITS 2 CHANNEL UNITS VOLUMETRIC UNITS Display HOLD LOW HIGH HHIGH OTHER YES NO 2 PATH ERROR HANDLING OF LCD PARAMS SUM DIF AVE CH1 CH2 TOTL TOTL TIME MDOT VEL VOLUM SS up SS do SNDSP T up MASS MASS REYN Tot K PEAK Qup Tdown DELTA AMPdn CNTup CNTdn P up Qdown AMPup PRESR Z AcVOL StVOL P dn TEMP Tsupr Rho Kelvn kPa Vinst Tsat QUAL until all channels set up appears only for 2 Chan...

Page 152: ...BEL RTD Inputs Low High C D A B FREQ OFF CH1 CH2 SUM DIF AVE Measurement Name Slot 1 Outputs TOTL TOTL TIME MDOT VEL VOLUM SS up SS do SNDSP T up MASS MASS REYN Tot K PEAK Qup Tdown DELTA AMPdn CNTup CNTdn P up Qdown AMPup PRESR Z AcVOL StVOL P dn TEMP Tsupr Rho Kelvn kPa Vinst Tsat QUAL BASE FULL Frequency Outputs FULL SCALE FREQUENCY C D A B TEMP PRESR Input measurement Analog Inputs SPEC OFF LA...

Page 153: ..._U Avg Mass Flow MASS_U Ch1 Fwd Mass Totals MTOT_U Ch2 Fwd Mass Totals MTOT_U Avg Fwd Mass Totals MTOT_U Ch1 Rev Mass Totals MTOT_U Ch2 Rev Mass Totals MTOT_U Avg Rev Mass Totals MTOT_U Ch1 Mass Tot Digits none Ch2 Mass Tot Digits none Avg Mass Tot Digits none Ch1 Timer sec Ch2 Timer sec Avg Timer sec Ch1 Error Code none Ch2 Error Code none Avg Error Code none Ch1 SSUP none Ch2 SSUP none Avg SSUP ...

Page 154: ...d in the OOS out of service mode before executing E 3 Selecting the Desired Measurements To set the measurement unit for each AI 1 Double click on the FLOW Transducer Block in the tree under GEFlow XGM 2 Select the Others tab and open the drop down list for the PRIMARY_SELECTOR and SECONDARY_SELECTOR refer to Figure 57 on page 145 3 Choose the unit from the list see Figure 57 on page 145 This unit...

Page 155: ... Selecting the Desired Measurements cont 4 After the desired measurements have been selected for the PRIMARY and SECONDARY SELECTOR choose the unit system UNIT_SELECTOR above the PRIMARY_SELECTOR that has been programmed in the flowmeter English or SI Figure 57 Primary Selector Drop Down List ...

Page 156: ...ou wish to set the units ANALOG_INPUT_1 or ANALOG_INPUT_2 in the tree under GEFlow XGM see Figure 56 on page 144 2 Select the Scaling tab and set the unit for the measurement based on the flowmeter settings For example if the flowmeter was set to use the metric unit system and the PRIMARY_SELECTOR was set to use VELOCITY you would choose m s for the unit as shown in Figure 58 Figure 58 Units Index...

Page 157: ...e click on the FLOW transducer block in the tree under GEFlow XGM see Figure 56 on page 144 2 Select the Others tab and scroll down to the CLEAR_TOTALIZERS listing 3 Select Clear from the drop down list box see Figure 59 4 After the totals have been reset select Normal from the drop down list box to resume total accumulation Figure 59 Clear Totalizers Drop Down List ...

Page 158: ...tup using the Function Block Application editor The flowmeter AI blocks along with the AO and PID of another device on the network are displayed We have connected the AI_1 OUT of the flowmeter to the CAS IN of the AO block We have also connected the AI_2 OUT of the flowmeter to the CAS IN of the PID block Figure 60 Function Block Application XGM ...

Page 159: ...e special programming software to configure a good status on the PV Yes 1 9 Does the Device feature a Non Volatile Memory protection feature If so what are the steps to disable enable from the Resource or Transducer block No 1 10 HOST System registered Yokogawa TBD Emerson Honeywell TBD ABB TBD 1 11 Design Concept Entity Concept 1 12 FISCO Compliant No 1 13 FNICO Compliant Yes models XGM868i XGS86...

Page 160: ...m Voltage 9 VDC 3 41 Device Resistance Ohms fieldbus terminal to 30 Mega Ohms increasing 3 42 Device Resistance Ohms fieldbus terminal to Ground open circuit 20 MOhm 3 43 Device Resistance Ohms fieldbus terminal to Ground open circuit 20 MOhm 3 51 Capacitance microF to 1 pF 3 52 Capacitance microF to Ground 634 7 pF 3 53 Capacitance microF to Ground 635 7 pF 3 6 4 wire Device if so what AC DC Volt...

Page 161: ...rsion of the FF spec FF 902 FS 1 4 7 2 Block Type Flow Display 7 3 Block Class Std Enhanced Custom Custom Custom Does the Device support Methods in the Resource and Transducer Blocks If it does then the functionality will be tested No 7 4 Special Features besides Methods multiple VIEWS etc Multiple View3s and View4s Device specific advanced diagnostics Yes 8 Function Blocks 8 1 Does the Device sup...

Page 162: ...short ton 1097 kilograms per cubic meter 1107 pounds per cubic foot 1142 pounds per square inch absolute 1143 pounds per square inch gauge 1324 kilogram per hour 1360 standard cubic feet per minute 1361 standard cubic feet per hour 1526 Standard cubic meter 20 1 atm 1547 kilopascal absolute 1548 kilopascal guage 1590 bar guage replaced with 34002 1597 Bar absolute replaced with 34001 1653 cubic Me...

Page 163: ...anged to 34026 33065 thousands of pounds changed to 34027 33066 mole weight changed to 34024 33067 thousands of kilograms removed 1322 kilogram per second 1323 kilogram per minute 1325 kilogram per day 1326 metric ton per second 1327 metric ton per minute 1328 metric ton per hour 1329 metric ton per day 1330 pound per second 1331 pound per minute 1332 pound per hour 1333 pound per day 1334 short t...

Page 164: ...ic meter per day 20 1atm 34000 Actual cubic meters 34001 Bar absolute 34002 bar gauge 34003 million pounds 34004 millions of actual cubic feet 34005 Millions of actual cubic feet per day 34006 Millions of actual cubic feet per hour 34007 Millions of actual cubic feet per minute 34008 Millions of actual cubic feet per second 34009 millions of actual cubic meters 34010 Millions of pounds per day 340...

Page 165: ... cubic feet per day 34030 Thousands of actual cubic feet per hour 34031 Thousands of actual cubic feet per minute 34032 Thousands of actual cubic feet per second 34033 Thousands of actual cubic meters 34034 Thousands of pounds per day 34035 Thousands of pounds per hour 34036 Thousands of pounds per minute 34037 Thousands of pounds per second 34038 Thousands of standard cubic feet 34039 Thousands o...

Page 166: ... of Standard cubic meters per hour 34046 Thousands of Standard cubic meters per minute 34047 Thousands of Standard cubic meters per second 9 3 Channel 2 Channel 2 Primary Value Secondary Value AI Block Units same as channel 1 Table 33 GE Fieldbus Device Capability XGX868 Family of Meter Types cont ...

Page 167: ...0 in PanaView 101 B Base Pressure Entering 13 104 in PanaView 104 Base Temperature Entering 12 103 Basic Programming Minimum Requirements 94 Burst Mode 16 108 C Calibration Factor in PanaView 101 Channel Activating 5 Entering Mass Flow Units 7 22 Entering Totalizer Units 6 Entering Volumetric Units 6 System Data 6 Channel Menu I O Submenu in PanaView 102 in PanaView 97 Set Up Submenu 106 Channel M...

Page 168: ...ting up with PanaView 89 F Fluid Type Entering 10 in PanaView 101 Foundation Fieldbus 36 143 Frequency Card Setting Up 123 Frequency Outputs Option Card 34 Setting Up 34 Frequency in PanaView 100 FWD Mass 117 FWD Total 117 G Global Entering Mass Flow Units 22 Entering System Data 21 Entering Totalizer Units 22 112 Entering Volumetric Units 21 Menu 20 Global Menu Comm Port Submenu 125 I O Submenu 1...

Page 169: ...ewing 66 Window 63 Low Pressure Switch Entering 13 Entering in PanaView 104 M M S Switch 16 108 Magnetic Keypad 3 MASS 26 Mass Flow 117 Activating 19 Activating in PanaView 110 Entering Channel Units 7 22 Entering Data in PanaView 99 Entering Global Units 22 Mass Flow Data in Global Menu PanaView 113 MDOT 26 Memory Clearing Log File 72 Clearing Site Data 71 Clearing Transmitter 71 Menus Edit 88 Ou...

Page 170: ... Wall Thickness 9 101 PIPE in PanaView 100 Plotting Signal 52 PRESR 27 Pressure Input Entering 12 Entering in PanaView 104 Pressure Units Selecting in PanaView 111 Printing Data 69 PROG Menu in Keypad Program 41 Properties Option in Edit Menu PanaView 132 Pulses Number of 108 Q Qdown 26 Quality Input 13 105 Qup 26 R R Window cycles 108 Requesting MODBUS Parameters 38 Response Time 108 Setting 17 R...

Page 171: ... Special Transducers Entering Number 8 SS do 26 SS up 26 Startup See Basic Programming 94 Static Density Activating in PanaView 110 Static Density Entering 19 Status of Logs 66 Status Submenu 97 Steam Input Type 11 102 StVOL 27 System Global Submenu 111 130 System Data Channel Entering 6 Entering in PanaView 98 Global Entering 21 Mass Flow Units Entering Channel 7 22 Mass Flow Units Entering Globa...

Page 172: ...CD Setting Up in PanaView 124 Tup 26 Two Path Error Handling 23 U Up Arrow Key 3 User Program Global Menu 110 I O Submenu113 116 125 Error Option114 LCD Option44 LCD Option124 System Submenu111 130 Minimum Requirements 94 Status Submenu 97 V V Averaging Option in PanaView 108 V Averaging Setting 17 VEL 26 Velocity 117 High Limit 15 Low Limit 15 Velocity High Limit 107 Velocity Low Limit 107 Viewin...

Page 173: ...ticular purpose and warranties arising from course of dealing or usage or trade Return Policy If a GE Sensing instrument malfunctions within the warranty period the following procedure must be completed 1 Notify GE Sensing giving full details of the problem and provide the model number and serial number of the instrument If the nature of the problem indicates the need for factory service GE Sensin...

Page 174: ...Warranty 164 DigitalFlow XGS868i Programming Manual no content intended for this page ...

Page 175: ......

Page 176: ...Certified Company www ge mcs com en about_us quality html Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S A Tel 800 833 9438 toll free 978 437 1000 E mail sensing ge com Ireland Sensing House Shannon Free Zone East Shannon County Clare Ireland Tel 353 0 61 470200 E mail gesensingsnnservices ge com ...

Reviews: