Appendix C. Programming the XGS868i Using PanaView™
108
DigitalFlow™ XGS868i Programming Manual
After responding to the above prompt, the meter returns to the
SET UP
option window.
C.5.5b The Default Setup Option
Use this option to initialize (reset) all of the parameters within the
Set Up
Signal
menu back to their default values.
Complete the following steps to reset all of the parameters:
1.
From the
Set Up
option menu, highlight and double-click on
Default Setup
in the center pane.
2.
Double-click on
No
to keep the current values or on
Yes
to reset all values to their default settings.
After responding to the above prompt, the meter returns to the
Set Up
option window.
C.5.5c The V Averaging Option
Use this option to specify the number of readings that occur before the meter will respond to a step change in flow rate.
In general, the smaller the number of readings, the less steady the display will appear. Complete the following steps to
set the response time:
1.
From the
Set Up
option menu, highlight and double-click on
V Averaging
in the center pane.
2.
Double-click on the desired
Response Time
(from 1 to 60 sec).
For best results, select the
STATS
(statistics) option, as this increases the response time under steady flow conditions
while still allowing a rapid response to changes in flow rate.
After responding to the above prompt, the meter returns to the
Set Up
option window.
M>S Switch
0 to 250
μ
sec
50
μ
sec
If the burst mode is set to Skan/Measure (S/M), the meter switches from
Skan to Measure Mode when Delta T is less than the M>S_Switch value.
DO NOT
change this value unless advised to do so by the factory.
# Shifts
0 to 10
3
The number of shifts corresponds to the actual number of transmits per
cycle (number of signals added together in one direction to produce an
averaged signal for one interrogation of the fluid) and need only be
changed if the environment is very noisy or the acoustic signal is weak.
A Divisor
0.1 to 10
2.5
A Divisor is used to calculate the Measure Mode integrated threshold
level and is not normally changed.
# Transmit
Pulses
1 to 16
4
# Transmit Pulses specifies the number of pulses in a burst. For difficult
conditions (i.e. long paths, high velocity or high temperature), settings as
high as 16 may be necessary.
T Window
(cycles)
0 to 1000
0
The XGS868i calculates the size of the transmit window based on pipe
size and fluid sound speed. However, for diagnostic purposes, the
window size may be reset.
R Window
(cycles)
10 to 128
10
The XGS868i calculates the size of the receive window based on pipe
size and fluid soundspeed. However, for diagnostic purposes, the
window size may be reset.
Table 20: Transducer Signal Settings (cont.)
Transducer
Signal
Parameters
Range
Default
Value
Description
Summary of Contents for DigitalFlow XGS868i
Page 2: ......
Page 4: ... no content intended for this page ...
Page 10: ...Contents viii DigitalFlow XGS868i Programming Manual ...
Page 52: ...Chapter 1 Programming Site Data 42 DigitalFlow XGS868i Programming Manual ...
Page 85: ...DigitalFlow XGS868i Programming Manual 75 Appendix A Menu Maps Appendix A Menu Maps ...
Page 86: ...76 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 96: ...86 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 146: ...Appendix C Programming the XGS868i Using PanaView 136 DigitalFlow XGS868i Programming Manual ...
Page 148: ...138 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 174: ...Warranty 164 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 175: ......