38
M1110150 006
© 2007 by General Electric Company. All rights reserved.
Chapter 9: Technical Reference
Table 4
Guidance and manufacturer’s declaration – electromagnetic emissions
The BiliSoft Phototherapy System is intended for use in the electromagnetic environment specified below.
The customer or the user of the BiliSoft Phototherapy System should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment- guidance
RF emissions CISPR 11
Group 1
The BiliSoft Phototherapy System uses
RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The BiliSoft Phototherapy System is suitable
for use in all establishments, including
domestic establishments and and those
directly connected to the public low voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies