36
M1110150 006
© 2007 by General Electric Company. All rights reserved.
Chapter 9: Technical Reference
Table 2
Guidance and manufacturer’s declaration - electromagnetic immunity
The BiliSoft Phototherapy System is intended for use in the electromagnetic environment specified below.
The customer or the user of the BiliSoft Phototherapy System should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test Level Compliance
level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 KHz
3 V/m
80 Mhz to 2.5 Ghz
3V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the BiliSoft Phototherapy Unit,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
150 kHz to 80 MHZ
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and
d
is
the recommended separation distance in
meters (m).
Field strengths for fixed RF transmitters
as determined by an electromagnetic site
survey (see Note 2 a) should be less than the
compliance level in each frequency range (see
Note 2 b).
Interference may occur in the vicinity of
equipment. Marked with the following symbol:
NOTE 1:
At 80 MHz to 800 MHz, the frequency range applies.