Section 15 - Calibrate Channels
37
15. Calibrate Channels
Before you begin Channel Calibration, you must choose the full scale range and set the
appropriate jumpers
(See Section 18.6)
. Calibrate both channels of your monitor. To
calibrate channels you will need these instruments and tools
•
multimeter
•
function generator
•
screwdriver
Use this procedure to calibrate both channels.
Calibrating a channel involves these main steps:
•
Set the filter option to No Filters
•
Connect and adjust the test instruments
•
Calibrate the channels
•
Prepare to bring the monitor on line