Section 22 - Appendix B - Monitor Options
53
22.5 Vibration Full Scale Range Options
To set the vibration full-scale range option, remove jumpers W15E-H, W24A-D, W25A-D.
Install jumpers specified in the following table.
Vibration Full Scale Range Options
Install Jumpers
Vibration Full Scale Range
W15 W24
W25
-01
-02
-03
-04
-05
-11
-12
-13
-14
-15
0 - 3 mil pp
0 - 5 mil pp
0 - 10 mil pp
0 - 15 mil pp
0 - 20 mil pp
0 - 100
µ
m pp
0 - 150
µ
m pp
0 - 200
µ
m pp
0 - 400
µ
m pp
0 - 500
µ
m pp
E, G
E, G
E, G
E, G
F, H
E, G
F, H
E, G
F, H
E, G
A
C
A, D
A, C, D
A, D
B
A
A, C
A, C
A, B, C, D
A
C
A, D
A, C, D
A, D
B
A
A, C
A, C
A, B, C, D
Replacement scales are included with the manual.
Caution
Changing full scale requires re-
calibration of both channels (see
Calibrate Channels section).