81
Adjusting the no breath time
If the monitor has the NeoResp license, in the neonatal mode, the
impedance respiration can set up the detect time for alarm “No
Breath”.
1. Press the
Others
key
2. Select
Resp Setup
-
No Breath Time
The default detection time is 15s.
Adjusting respiration alarms
1. Press the
Others
key.
2. Select
Resp Setup
-
Resp Rate Alarm
.
• To set up the limits, select
Adjust Limits
.
• To OFF/ON the limits, select
RR Alarm
.
3. You can also select
Alarms Setup
to adjust RESP or related
alarms.
Troubleshooting
• Measurement fails:
−
Check the electrode quality and positioning.
−
Adjust the detection limits. During ventilator supported
breathing, the respiration calculation may count only
ventilator-produced inspirations and expirations.
−
Other electrical devices may interfere with the measurement.